Download PDF
Review  |  Open Access  |  11 Sep 2025

Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

Views: 38 |  Downloads: 3 |  Cited:  0
Extracell Vesicles Circ Nucleic Acids. 2025;6:560-79.
10.20517/evcna.2025.10 |  © The Author(s) 2025.
Author Information
Article Notes
Cite This Article

Abstract

In the past decade, extracellular vesicles (EVs) have gained increasing attention in biomedical research. These membrane-bound particles are naturally secreted by cells under both physiological and pathological conditions, and they exhibit a wide range of sizes and molecular compositions. EVs transport bioactive molecules - such as proteins, nucleic acids, and lipids - making them ideal candidates for biomarker discovery. Consequently, their accurate characterization and quantification are critical for understanding their roles in intercellular communication and evaluating their potential in diagnostics, prognostics, disease monitoring, and therapeutic applications. Microfluidic technologies offer promising solutions for EV analysis, addressing key limitations of conventional methods by enabling precise and sensitive measurements with small sample volumes. While microfluidic devices have been predominantly used for EV separation and isolation, their application in EV quantification remains underexplored. Compared to traditional techniques like nanoparticle tracking analysis or flow cytometry, microfluidic systems can provide faster, more accessible alternatives for EV quantification. This review summarizes recent advances in microfluidic technologies for EV quantification, discussing their advantages, current limitations, and future prospects.

Keywords

Extracellular vesicles, microfluidics, on-chip quantification, molecular biomarkers, lab-on-a-chip, integrated sensing

INTRODUCTION

Intracellular compartmentalization of chemical processes within membrane-enclosed reaction vessels is essential to eukaryotic cells, and nearly all cells release small vesicles into their surroundings. Although extracellular vesicles (EVs) have been known for over half a century[1-3], recent research has revealed their crucial roles in intercellular communication, particularly their therapeutic and diagnostic potential[4]. Despite the potential benefits of using EVs as biomarkers, their clinical application faces challenges in standardization and reproducibility of isolation, detection, and quantification, largely due to their small size (typically 50-1,000 nm), abundance, and physicochemical properties[5]. EVs are typically classified based on their origin [Figure 1A] or size [Figure 1B][6], given the preponderance of size in isolation techniques (e.g., centrifugation and chromatography-based approaches). Their structure is known to replicate many (but possibly not all) of the features of their cells of origin[7], carrying with them various biomarkers, both internally and at the surface [Figure 1C-E]. Of note, the concentration of tetraspanins in EV membranes can exceed that of their cell of origin[8], while lipidic composition is also distinct from cells, with a dense proteoglycan shell surrounding the membrane[9], with these characteristics being thus explored for both EV isolation and quantification. EVs also have distinct mechanical properties [Figure 1F], essential for gold-standard isolation methods based on centrifugation, and optical properties [Figure 1G], which are exploited by typical technologies for EV quantification and characterization (e.g., nanoparticle tracking analysis - NTA)[6,10].

Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

Figure 1. Main features and characteristics of EVs and microfluidic systems. Origin (A) and size range (B) of EVs. (C) Representative structure of a 100 nm EV, compared to the sizes of its typical molecular components. Main surface (D), molecular (E), mechanical (F), and optical (G) characteristics of EVs. Key features of microfluidic systems and chips (H), and highlighted approaches for microfluidic EV quantification (I). EVs: Extracellular vesicles; SERS: surface-enhanced raman spectroscopy.

However, the standardization of technologies and procedures to ensure comparable and reproducible results remains one of the major challenges in EV-related research. To address this, the International Society of Extracellular Vesicles (ISEV) provides recommendations and guidelines in this field[6], not only specifying the minimum EV sample characterization required, but also advocating for consistent nomenclature. Regarding the latter, the use of the term EV is recommended when specific markers of biogenesis pathways are not identified. Researchers are thus encouraged to describe EVs according to other parameters, such as size, biochemical composition, or characteristics related to their cells of origin. In this review, the terms “exosome” and “EV” were used in accordance with the terminology employed by the authors of the works discussed herein.

The widespread use of current “gold-standard” methods for EV quantification, such as NTA, results in well-established protocol guidelines[6]. NTA is typically used to characterize particle size and concentration. However, in complex matrices like human samples (i.e., blood, plasma, serum), the presence of non-vesicular extracellular particles significantly complicates NTA analyses. Consequently, NTA is often unsuitable for clinical applications due to its cumbersome use and high upfront cost[5]. In contrast, while microfluidic platforms are in an earlier stage of development and implementation than NTA, they offer greater flexibility in the analytical information that can be obtained from EVs, including cargo and protein analysis, nucleic acids, and complex omic studies.

This advantage drives the development of seamless EV quantification technologies. Microfluidics enables the integration, miniaturization, and automation of complex biochemical analyses, leveraging the small feature sizes of microfluidic systems (from 100s of nm to a few mm - Figure 1H)[11,12]. The design and fabrication flexibility of microfluidics allows for the development of systems that manipulate microscopic objects while reducing material costs. Devices can be designed with large surface-to-volume ratios, enhancing EV absorption, and can be tailored for specific EV capture schemes[13,14]. Furthermore, the integration of external/internal sensing approaches enables the analysis of individual captured particles. This flexibility also allows for high-throughput processing of both small and large volumes of complex samples, a critical consideration for clinical applications. Thus, the strong synergy between EVs and microfluidics is evident. At the intersection of these two fields, a major focus has been on the isolation of EVs, exploring how phenomena at the microscale could enhance this process[13-16]. Moreover, applying microfluidic devices for the analysis and characterization of EVs and their cargo/biomarkers has become a key area of interest in recent years[13-18]. Other EV-related efforts have been utilized in a variety of fields, including regenerative medicine[19] or liquid biopsy and oncology[20-22], for example. Within these broader trends, there is an increasing interest in not only performing the on-chip isolation of EVs, but also in integrating the capability to perform on-chip quantification, analysis, and characterization of isolated EVs. This review provides a concise summary of recent advances and current trends in microfluidics-based EV quantification technologies, covering both surface marker-specific and -independent approaches, as well as the quantification of EV internal cargo - Figure 1I.

SURFACE MARKER-SPECIFIC EV QUANTIFICATION

EV surface proteins can be targeted for quantification, characterization, functional analysis, and biomarker studies. Transmembrane proteins, such as the commonly found tetraspanins CD63, CD9, and CD81, are often used, and pathology-specific markers can also be targeted for EV quantification. The main features of surface marker-specific quantification platforms are summarized in Table 1.

Table 1

Summary of techniques developed for microfluidic on-chip quantification based on surface markers

Authors Type Principle Performance parameters Sample type Sample pre-processing Markers targeted
Chen et al.[28] Fluorescence Membrane EV isolation/counting microfluidic platform LOD: 105 particles/mL (108 for clinical samples)
Range: 1 × 105-4 × 106 particles/mL
Volume: 2 µL
Whole blood N CD63
Chen et al.[27] Fluorescence Membrane-based filtration + magnetic-bead based immunoassay LOD: 3 × 1010 particles/mL
Linear range: 1-6 × 1012 particles/mL
Whole blood N CD63
Yang et al.[24] Fluorescence Droplet-based optofluidic platform LOD: 9 × 103 particles/mL FBS spiked with neuronal EVs N CD81
Liu et al.[29] Fluorescence Droplet-based single-exosome counting immunoassay LOD: 104 particles/mL
Range: 104-108 particles/mL
Human serum N CD63 to capture and GPC-1 to detect
Wu et al.[31] Fluorescence Fluorescence immunoassay on a herringbone structure LOD (based on protein concentration): 9.95 ng/µL Mice plasma N CD63 and PD-L1
Walker et al.[30] Fluorescence Fluorescently labeled EVs captured in a NPN membrane Analyzes up to 104 EVs
Selective detection of tumor-related Evs - 1%
Human plasma
CM
Labeled and prefiltered CD9 and ICAM-1
Ji et al.[33] Fluorescence Magnetic EV isolation, Acoustic wave-assisted mixing; Monolayer-fluorescence counting using a step-wedge LOD: 8.5 × 102 particles/mL CM
Saliva
N EpCAM
Zhao et al.[23] Fluorescence DEP trapping in microwells LOD: 193 particles/mL
Linear range: 1.4 × 103-1.4 × 108 particles /mL
CM UC CD63 to trap and to detect: CD81, CEA, EpCAM, CD147, and AFP
Lee et al.[25] Fluorescence Single EV analysis with multiplexed immunofluorescence Simultaneous analysis of > 103 EV
Detection of 11 markers
CM Differential centrifugation, filtration CD9, CD63, CD81, EGFR, EGFRvIII, IDH1, IDH1R132, PDPN, PDGFRα, PD-L1, PD-L2
Zhou et al.[32] Chemiluminescence Inertial separation + capture on AuNPs + chemiluminescence LOD: 9.5 × 104 particles/mL
Range: 2.5 × 105-2.5 × 1011 particles/mL
Whole blood N CD24
CD81, EpCAM
Wang et al.[34] Colorimetric Centrifugal chip + capture on magnetic NPs LOD: 106 particles/mL
Volume: 100 μL of blood
Time: 10 min
CM
Plasma (human)
Blood (mice)
UC CD63 to capture
CEA, CA125, EGFR
Vaidyanathan et al.[35] Colorimetric Capture with AC-EHDs + Catalytic oxidation of HRP with TMB LOD: 2.8 × 106 particles/mL CM Exosome isolation reagent CD81 and HER2
Stollmann et al.[36] Digital in-line holography Optofluidics Range: 1.9 × 108 to 7.6 × 109 particles/mL
Throughput: 104 events/h
CM Ultrafiltration and SEC CD63, CD9, CD81, EpCAM, CA-125, HE-4
Cavallaro et al.[38] Electrical Electrical current in a capillary LOD: 2.8 × 108 particles/mL CM SEC or UC CD63, CD9, EGFR
Kim et al.[40] Electrical Inertial and DEP separation + GO biosensor Linear range: 104 to 106 particles/mL CM N CD63
Talebian Gevari et al.[39] Electrical Streaming current biosensing LOD: 104 particles/mL
Size range 30-200 nm
CM
Plasma
SEC CD9, CD81, CD73, and PD-L1
Li et al.[42] Electrochemical Electrochemical microfluidic aptasensor LOD: 1.4 × 104 particles/mL
Linear range: 105-109 particles/mL
Time: 60 min
CM
Plasma
Exosome extraction Kit EpCAM aptamer
Wang et al.[41] Electrochemical Dual microfluidic filtration chip + four SPEs LOD: 1 × 104 particles/mL
Time: 60 min
Whole blood N PMSA, EGFR, CD81, and CEA
Qian et al.[43] Magnetic DNA-mediated magnetic detection LOD: 1.98 × 103 particles/mL
Linear Range: 103-107 particles/mL
Detection time: ~1 h
CM
Simulated serum
UC and filtration CD9, CD63
Ho et al.[46] SERS Droplet microfluidics + SERS LOD: 104.5 particles/mL
Time: 5 min
CM
Plasma
N HER2, CD44, CD63
Wang et al.[44] SERS SERS microfluidic biosensor LOD (in particles/mL): MCSP/ MCAM - 1 × 103, CD61/CD63 - 1 × 105
Volume: 5 µL
CM
Plasma
CM - SEC + UF
Plasma - tenfold dilution
MCSP, MCAM, CD61, CD63
Zhou et al.[45] SERS SERS microfluidic biosensor LOD: 106 particles/mL for LacdiNAc and T antigen and 105 particles/mL for CD81 CM
Plasma
SEC and UC CD81, LacdiNAc, T antigen
Hao et al.[47] SERS Acoustic enrichment + immunofluorescent/SERS bimodal biosensors Fluorescence mode - LOD: 1.3 × 106 particles/mL
SERS mode - LOD: 2 × 104 particles/mL
Linear range: 105-1011 particles/mL
Commercial solution (EV standard)
Plasma
Acoustic isolation for plasma samples CD63

Fluorescence

Many microfluidic techniques for surface biomarker-based EV quantification use fluorescent labeling to measure the intensity of captured vesicles. This enables a correlation between biomarker abundance and fluorescence signal intensity. Additionally, when targeting common EV markers (e.g., tetraspanins), these methods provide an estimate of the total EV concentration.

For example, a system using anti-CD63 labeling and samples from A549 cell culture supernatant achieved a limit of detection (LOD) of 193 exosomes/mL, with the relative abundance of other surface proteins (CD81, CD147, EpCAM, CEA, and AFP) also being assessed[23]. A droplet-based microfluidic platform was shown to provide information from single CD81+ EVs from human induced pluripotent stem cells differentiated to neurons (LOD: 9 × 103 EVs/mL)[24]. Other systems have been optimized for single vesicle detection and analysis of multiple surface markers by immunofluorescence using previously isolated EVs from human glioblastoma cell lines[25]. Another system simultaneously assessed EV sample purity, particle size, and EV quantities with CD63+ EVs isolated from a mesenchymal stem cell line within a range of 1.25-5 × 1013 particles/mL[26].

Some microfluidic platforms can use whole blood directly by integrating a filtration module to obtain platelet-free plasma. For example, circulating EVs can be captured using an ELISA-based method, with EV quantification based on the fluorescence intensity of CD63-labeled EVs[27,28]. With only 2 µL of whole blood, the system achieved a LOD of 105 CD63+ EVs/mL[28] - Figure 2A. In a different approach, a droplet microfluidics system used magnetic beads for EV capture, followed by quantification of CD63+GPC-1+ exosomes from breast cancer patient serum (LOD: 104 exosomes/mL)[29]. A microfluidic system that showed good results for counting and characterizing EVs from both cell cultures and plasma samples with minimal handling has also been developed[30] - Figure 2B. Labeled EVs with CFSE were detected in a linear range of 102-106 EVs (from cell cultures) and 102-104 EVs (from plasma). For further characterization, colocalization studies for the surface markers CD9, ICAM-1, PD-1, and PD-L1 were also conducted[30]. Another interesting strategy was developed to monitor nascent EVs upon the administration of an immunotherapy[31]. The combination of metabolic labeling of EVs and detection of PD-L1+ and PD-L1+CD63+ plasma-derived EVs that were secreted over time in response to an anti-PD-L1 treatment in a mouse model was performed. With the metabolic labeling approach, the chip achieved a LOD of 9.95 ng/µL (quantified by the bicinchoninic acid protein assay). Chemiluminescence was also explored to quantify CD63+CD81+ EVs (LOD: 9.5 × 104 EVs/mL), as well as the levels of CD24+ and EpCAM+ EVs in cell lines and in plasma samples from patients with ovarian cancer and healthy controls[32].

Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

Figure 2. Fluorescence-based approaches for surface marker-based EV quantification. (A) ELISA-based EV quantification from blood samples (adapted from ref.[27]); (B) Detection of fluorescent EVs trapped on ultrathin nanopore membrane (adapted from ref.[30]). EVs: Extracellular vesicles; ELISA: enzyme-linked immunosorbent assay; HRP: horseradish peroxidase; NPN: nanoporous silicon nitride; FOV: field of view.

This family of techniques has also been extended to other biofluids. Saliva-derived EVs expressing EpCAM were detected in a platform where the sandwich complexes formed between magnetic beads, fluorescent beads, and EpCAM+ EVs were arranged in a monolayer and counted individually. This method achieved a LOD of 850 particles/mL and was tested in samples from healthy controls and oral cancer patients[33].

Colorimetric and holography

Instead of fluorescence emission, colorimetric changes can also quantify EVs within a device. A centrifugal disk system capable of efficient EV capturing (using CD63-targeting carboxyl magnetic beads) was integrated with EV labeling (antibodies targeting CEA, CA125, and EGFR) for on-chip detection based on a horseradish peroxidase-associated secondary antibody[34] - Figure 3A. The system handled small volumes (100 μL) of blood samples, providing visual results within 10 min with a LOD of 1 × 106 particles/mL. The Trau group used an electrohydrodynamic device to create fluid flow a few nanometers away from an electrode surface; the induced surface shear forces (“nanoshearing”) improved immuno-based EV capture at the electrode surface[35]. Using the catalytic oxidation of horseradish peroxidase, a rapid naked-eye colorimetric assessment of EV capture is possible, as well as UV-visible spectroscopy measurements, for the analysis of multiple EV biomarkers (e.g., HER2 and PSA) with a LOD of 2.8 × 106 EVs/mL.

Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

Figure 3. Colorimetric and holography-based approaches for surface marker-based EV quantification. (A) Colorimetric EV quantification in centrifugal microfluidic disk (adapted from ref.[34]); (B) Digital in-line holography device for EV detection (adapted from ref.[36]). EVs: Extracellular vesicles; HRP: horseradish peroxidase; TMB: 3,3’,5,5’-tetramethylbenzidine.

Digital in-line holography can also be integrated for on-chip EV quantification, providing high-throughput molecular fingerprint - Figure 3B[36]. This optofluidic system contained a liposome-covered substrate, treated with a biotin-based protocol for antibody-based EV capture. By targeting EV-specific and cancer-specific biomarkers for four different ovarian cancer subtypes, the system could identify the presence of EVs with concentrations above 1 × 108 particles/mL, creating a unique fingerprint for each cancer subtype.

Electrochemical and magnetic sensors

Electrochemical and magnetic sensing are alternative strategies for surface marker-specific EV quantification. These methods use affinity-based detection mechanisms to capture and analyze EVs via electrochemical or magnetic signal transduction. Electrochemical techniques offer high-sensitivity detection by measuring changes in electrical properties upon EV binding, while magnetic sensors use nanoparticle-based labeling for enhanced signal amplification. These sensors often provide higher sensitivity than fluorescence methods and can be used with portable readers, eliminating the need for large equipment like the fluorescence microscopes required for most fluorescence-based methods.

Cavallaro et al. targeted EGFR and CD63, using an electrokinetic sensor. In this device, changes in the streaming current produced by the affinity-based binding of EVs to the walls of a microcapillary were used for quantification (LOD of 2.8 × 108 particles/mL)[37]. This device was subsequently adapted for simultaneous detection of multiple EV surface markers[38]. In a follow-up study, it was functionalized with silane-PEG-biotin to detect PD-L1+ and CD73+ EVs (LOD of 1.4 × 104 particles/mL), including detection in plasma samples from a lung cancer patient[39]. Dielectrophoretic (DEP) forces, generated within an Ag-doped microfluidic device, and inertial effects have also been used for EV separation (83% efficiency), followed by EV detection using a CD63-based reduced graphene oxide-based biosensor (linear range of 104 to 106 particles/mL)[40]. Another approach employed antibody-coated screen-printed electrodes to capture isolated EVs, followed by the adsorption of metal-organic frameworks onto the EV surface, leading to the amplification of electrical signals[41]. The system enabled the detection of EV-associated tumor biomarkers and the classification of various breast cancer mouse models and clinical samples (LOD: 1 × 104 particles/mL) - Figure 4A. An integrated electrochemical microfluidic aptasensor, based on impedance spectroscopy, was shown to identify differences between EVs from lung cancer and healthy controls[42]. Using herringbone structures to enhance EpCAM-based EV capture onto the gold electrode, a biomolecular layer was formed after adsorption, causing an increase in impedance proportional to EV concentration (LOD: 1.4 × 104 particles/mL). A microfluidic magnetic detection system was also used to quantify EVs[43]. EVs were captured through a sandwich assay using a DNA tetrahedral-structured probe connected to streptavidin and biotin-anti-CD9, followed by labeling using CD63 aptamer-modified Fe3O4 magnetic nanoparticles. An induction coil-based magnetic detector was then able to measure the EV-associated magnetic signal (LOD: 1.98 × 103 particles/mL).

Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

Figure 4. Integrated sensors and SERS approaches for surface marker-based EV quantification. (A) Electrochemical EV quantification (adapted from ref.[41]); (B) Droplet microfluidics with integrated SERS for EV detection (adapted from ref.[46]). EVs: Extracellular vesicles; SERS: surface-enhanced raman spectroscopy.

Surface-enhanced raman spectroscopy

Recent works have explored the use of surface-enhanced raman spectroscopy (SERS) systems for EV quantification. In one such study, the Trau group employed a nanoshearing device to capture and quantify EVs based on melanoma-associated biomarkers: MCSP and MCAM (LOD: 1 × 103 particles/mL), and CD61 and CD63 (LOD: 1 × 105 particles/mL)[44]. Subsequently, SERS signatures of the captured EVs were obtained from a cohort of 20 early-stage melanoma patients and 21 healthy controls. A mean fold increase in SERS signal intensity ranging from 3.7 to 4.2 was observed in patients compared to controls. The nanoshearing system was later adapted for multiplex profiling of glycans in EV membranes from non-small cell lung cancer patients[45]. Two different glycans (LacdiNAc and T antigen), plus CD81, were used to label SERS nanotags for EV characterization. SERS signatures from EV glycans were assessed for 40 patients (LOD: 1 × 106 particles/mL) and revealed different glycopatterns between late-stage patients and healthy individuals, and between early-stage patients and patients with benign lung diseases.

A droplet microfluidics device was proposed to co-encapsulate EVs with gold nanoparticles coated by HER2-targeting aptamers, along with salt and Raman tags, for SERS detection[46] - Figure 4B. The higher affinity between aptamers and HER2-positive EVs, together with the high-salt conditions, leads to nanoparticle aggregation and generation of strong SERS signals for identification of HER2+ EVs from breast tumor cells (LOD: 1 × 104.5 particles/mL). A bimodal sensing approach, combining SERS with fluorescence, was also created[47]. The system used interdigitated transducers to generate acoustofluidic forces to control exosomes captured by CD63-aptamer-conjugated nanoparticles within a glass capillary. In fluorescent mode, fluorescently-labeled exosomes were concentrated at the center of the capillary for fluorescence quantification (LOD: 1.3 × 106 particles/mL), while in SERS mode, by changing the acoustofluidic frequency, the EV-coated nanoparticles were re-focused to the wall of the capillary, previously coated with nanorods for enhanced SERS signaling, for EV detection from human plasma samples (LOD: 2 × 104 particles/mL).

SERS-based methods offer highly sensitive detection and detailed molecular fingerprinting of EVs. However, they require complex plasmonic substrates and careful analysis to ensure batch-to-batch reproducibility of the nanostructures. This makes SERS ideal for applications requiring high precision and characterization, but less suitable for cost-effective, portable, and rapid point-of-care (PoC) testing. For applications where characterization, ultrasensitivity, and high specificity are critical, SERS is a powerful tool, whereas biosensors (such as electrochemical sensors) provide a more practical alternative when affordability, simplicity, and rapid detection are needed.

SURFACE MARKER-INDEPENDENT EV QUANTIFICATION

Surface marker-independent EV quantification methods offer an alternative approach for detecting and analyzing EVs based on their physical and chemical properties. These techniques eliminate specific labeling protocols, helping to preserve the native state of EVs and reduce potential biases. They often provide additional information, such as size distribution and surface charge, valuable for functional analyses and biomarker discovery. However, these methods also present challenges, particularly in distinguishing EVs from other particles of similar sizes. A summary of the main features of surface marker-independent quantification platforms is provided in Table 2.

Table 2

Summary of surface marker-independent on-chip quantification techniques (nonspecific labeling & label-free), and techniques for internal cargo quantification

Authors Type Principle Performance parameters Sample type Sample pre-processing
Paganini et al.[26] Fluorescence Hydrodynamic focusing + Fluorescence-based diffusion sizing Size range: nanometers to hundreds of nanometers
Range: 1010 to 1015 particles/mL
Volume: 2-4 µL
Completion time in minutes
Differentiates EVs from co-isolated contaminants
CM UC and UF
Friedrich et al.[48] Fluorescence Nanofluidic-based fluorescence flow cytometry Size range: down to 100 nm
Sensitivity: 170 fM to 500 pM vesicle concentration
Time: < 2 min
Volume: 20 μL sample volume
CM UC
Hong et al.[51] Fluorescence Opto-thermo-electrohydrodynamic tweezer + fluorescence Size: 30 and 150 nm
Range: down to 1.8 × 105 particles/mL
Commercial solution (EV standard) Not applicable (EV standard)
Gustafson et al.[49] Fluorescence DEP collection + fluorescence quantification Not provided Plasma N
Ware et al.[50] Fluorescence DEP collection + fluorescence quantification LOD: 4.31 × 109 BMVs/mL in plasma
Linear range: 2.8 × 109 to 7 × 109 BMVs/mL
Reproducibility: CV = 0.74
Volume: 25 µL
Plasma N
Kim et al.[52] Electrical Resistive pulse sensing High sensitivity (error of < 10%)
Throughput: > 200,000 EVs/s
Lower size limit: 50-100 nm
CM UC
Young et al.[53] Electrical Resistive pulse sensing 60 to 160 nm
Particle frequencies up to 1,000 per min
Allows for measuring zeta potential
CM
Milk
UC
Calado et al.[54] Electrical Resistive pulse sensing Size: down to 80-100 nm
Range: 106-1010 particles/mL
Throughput: 80 nL/s with particle concentrations up to 109/mL
CM
Human serum
UC
Cimorelli et al.[55] Electrical Resistive pulse sensing Size: 65-75 nm up to 2 μm
Range: 104-1012 particles/mL
CM
Urine
Plasma
Minimal (centrifugation and collected supernatant + dilution)
Jalali et al.[56] SERS Nanostructured device with plasmonic nanobowties LOD: 1.32 × 10⁵ particles/mL
Differentiation between glioblastoma EVs, non-cancerous EVs, and synthetic liposomes
CM Filtration and SEC
Cheng et al.[60] Cargo FET-based sensing Sensitivity: miR-21: LOD of 6.07 fM/miR-126: LOD of 23.82 fM CM
Plasma
UC
Sung et al.[59] Cargo Digital PCR quantification LOD: 11 copies/mL for miRNA-21
Accuracy: quantification accuracy of > 88%
Time: EV isolation: 4 h; miRNA extraction: 20 min
Plasma Magnetic immunoassay and lysis
Zhang et al.[61] Cargo PCR-free digital bioassay LOD: 18 aM for GAPDH mRNA
Specificity: discrimination between EWS-FLI1 fusion transcript variants
Efficiency: GAPDH mRNA: 43.5-64.6 copies/105 EVs. EWS-FLI1 mRNA: 0.277-6.5 copies/105 EVs
CM UC and lysis
Ramshani et al.[58] Cargo SAW-based lysing chip and IEM-based sensing LOD: 1 pM for miR-21
Quantification accuracy: < 10% uncertainty for miRNA levels
Specificity: differentiated between free-floating and EV-associated miRNAs
Plasma N
Li et al.[57] Cargo Electrochemical sensing with a sandwich immunosensor Sensitivity: 1 pg/mL for L1CAM
Time: 1.5 h
Sample Volume: 300 μL serum
Specificity: high specificity demonstrated against interference proteins like IgG, CD81, HSA, and IL-6
Serum N

Nonspecific labeling

These techniques use fluorescent nonspecific lipid dyes to stain the lipid bilayer of EVs for fluorescence-based quantification/analysis. This is the case of a nanofluidic flow cytometer for single lipid vesicles and EVs that enabled quantification and size distribution analysis as fluorescently labeled vesicles flow through parallel nanochannels under pressure-driven flow[48]. Single-particle analysis was performed by recording fluorescence signals, achieving detection of EVs down to 100 nm and quantification in the range of 170 fM to 500 pM. A different approach was a device that quantified fluorescent EVs while measuring the size distributions from their diffusion profiles[26]. This approach can detect EVs in a wide size range, in concentrations from 1010 to 1013 particles/mL, and differentiate EVs from co-isolated contaminants. Other groups have incorporated external fields into devices. The Ibsen group[49,50] developed a label-free EV quantification technique in blood plasma by employing DEP for EV collection and automating fluorescence quantification. This approach was applied to quantify bacterial membrane vesicles (BMVs) in the range of 2.8-7 × 109 BMVs/mL[50] - Figure 5A. These techniques were only validated with fluorescent immunostaining, but, in principle, can be easily turned into label-free techniques using nonspecific labels. A similar approach used opto-thermo-electrohydrodynamic tweezers for trapping and manipulation of single exosomes (30-150 nm)[51].

Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

Figure 5. Surface marker-independent EV quantification and EV cargo quantification. (A) DEP-based EV capturing and quantification (adapted from ref.[50]); (B) Nanofluidic resistive pulse sensing of EVs (adapted from ref.[54]); (C) L1CAM quantification from L1CAM-positive EVs (adapted from ref.[57]). EVs: Extracellular vesicles; DEP: dielectrophoretic; BMV: bacterial membrane vesicles; RPS: resistive pulse sensing; NTA: nanoparticle tracking analysis.

Label-free

Label-free techniques eliminate specific labeling protocols by exploiting the physical or intrinsic chemical properties of EVs. This helps preserve their native state and reduces potential biases, often providing additional information like size distribution. However, the size and heterogeneity of EVs, and the presence of similar particles in samples, make label-free technique development challenging, requiring robust designs to ensure specificity.

Resistive pulse sensing (RPS) is a well-established EV quantification approach. Based on resistive signals generated when particles traverse a nanopore, RPS has been integrated into microfluidic devices in several configurations. A multi-gate design was proposed to enhance throughput, including a sheathless focusing detection gate with three nanochannels and a reference gate for noise reduction[52]. A different approach used a multi-pore nanochannel device that also enabled the determination of the zeta-potential of the EVs[53]. More recently, a RPS chip was developed with two parallel microchannels connected by in-plane nanopores[54] - Figure 5B. This approach enabled calibration-free size distribution measurements across a dynamic range of 106 to 1010 particles/mL, but provided only orders of magnitude estimates for the concentration. Microfluidic RPS has even been commercialized, as exemplified by Spectradyne LLC (Torrance, CA, USA), which offers a system with microfluidic cartridges containing a nanoconstriction. This platform has been used to optimize a standardized procedure for measuring EV concentration and size distribution in unpurified biofluid samples from various sources[55]. While RPS systems excel in scalability, cost-effectiveness, and adaptability for high-throughput applications, their performance is constrained by the nanopore size. Furthermore, these devices are susceptible to clogging and often suffer from reproducibility limitations. The fabrication of nanopores also presents challenges, being complex, time-consuming, and costly.

An additional strategy was to use SERS by embedding the nanostructures in the microchannels. For instance, plasmonic nanobowties were used to amplify EV Raman signals and detect and analyze EVs in the samples, achieving a LOD of 1.32 × 105 particles/mL[56].

EV INTERNAL CARGO QUANTIFICATION

Microfluidic systems streamline the quantification of EV cargo biomarkers by integrating on-chip isolation, lysis, and analysis. One notable approach was a microfluidic electrochemical sandwich immunosensor [Figure 5C] for detecting L1CAM, a proposed biomarker for Parkinson’s disease. The system used immunomagnetic beads for EV isolation, on-chip lysis, and electrochemical protein detection, achieving a sensitivity of 1 pg/mL for L1CAM, with high specificity against potential interference proteins such as IgG, CD81, HSA, and IL-6[57].

MicroRNAs (miRNAs) are another class of biomarkers commonly analyzed. For example, a microfluidic platform incorporating surface acoustic waves for EV lysis and ion-exchange membranes for miRNA concentration and detection was developed[58]. This system quantified miR-21 with a detection limit of 1 pM and distinguished between free-floating and EV-associated miRNAs, enabling precise analysis of plasma samples. The Gwo-Bin Lee group[59,60] advanced automated microfluidic systems for miRNA quantification by combining CD63 antibody-coated magnetic beads for EV isolation with probe-coated magnetic beads for miRNA extraction from EV lysates. Using these systems, the group developed two quantification modules. A digital PCR module enabled miR-21 quantification with a detection limit of 11 copies/mL in ovarian cancer samples. A second approach used sensors based on field-effect transistors, achieving detection limits of 6.07 fM for miR-21 and 23.82 fM for miR-126, highlighting their versatility in detecting disease-related miRNAs. For mRNA analysis, a microwell-patterned microfluidic device for highly sensitive digital bioassays enabled absolute quantification of mRNA biomarkers from EV lysates using dual-probe hybridization and enzymatic signal amplification[61]. Using dual-probe hybridization and enzymatic signal amplification, the platform enabled absolute quantification of mRNA biomarkers from EV lysates. However, unlike other platforms, this system requires off-chip EV isolation and lysis prior to on-chip analysis.

DISCUSSION AND FUTURE PERSPECTIVES

EVs hold significant promise as biomarkers for diagnosis, prognosis, and treatment monitoring, as well as in drug delivery systems. Accurate quantification and characterization are therefore crucial to realizing their potential in these applications. However, current techniques for EV analysis are often costly and require specialized equipment and skilled operators. Microfluidic technology offers a potentially more accessible alternative. While most of the on-chip platforms focused primarily on EV capture and isolation, recent scientific exploration has increasingly focused on the development of microfluidic technologies for EV quantification and characterization. In this work, we have reviewed the most recent advances in this field. Most of the systems discussed are based on the presence of surface markers, either those commonly found on EVs (e.g., CD9, CD63, CD81), or disease-specific proteins of interest (e.g., HER2, CD24, EpCAM, PD-1, PD-L1). The measurement of the fluorescent intensity when EVs were captured was the most common approach. Electrochemical and SERS-based systems have also shown promise for detailed EV subpopulation studies relevant to clinical diagnosis. Some of these works employed droplet microfluidics[24,29,46], a branch of microfluidics dealing with segmented flow that is very well known for its application in the analysis of single cells. Researchers in both fields have adapted those technologies to EV analysis, enabling subpopulation studies of EVs expressing specific target proteins, which may contribute to unlocking the potential of EVs in clinical diagnosis. Droplet microfluidics offers significant advantages for EV analysis, including high-throughput capabilities and precise microenvironment control. For instance, simultaneous detection of membrane proteins and mRNA at the single EV level was achieved with high fidelity[62,63]. Similarly, Tong et al.[64] developed an all-in-one microfluidic workstation for high-throughput EV miRNA detection in non-small cell lung cancer. Further, Reynolds et al.[65] presented a double digital assay for single EV and single-molecule detection, again showing the potential for high-sensitivity analysis of individual EVs and their molecular cargo using microdroplet-enabled strategies. In addition, nonspecific labeling and label-free methods enable the quantification of intact EVs, suitable for subsequent functional studies or heterogeneous EV populations. Quantification of EV internal cargo, such as proteins and nucleic acids, has also been explored.

Comparing these strategies reveals that each has unique strengths and limitations, and that the choice between these approaches often depends on the intended application. Surface marker-independent approaches offer the advantage of preserving the native state of EVs, minimizing labeling biases, and enabling simultaneous analysis of physical properties. RPS integrated into high-throughput microfluidic devices offers scalability and cost-effectiveness. However, distinguishing EVs from similarly sized particles in complex samples and addressing EV population heterogeneity remain challenges. In contrast, surface marker-specific quantification techniques leverage EV membrane biomarkers for detection and quantification. Fluorescence-based assays, for example, often utilize fluorophore-conjugated antibodies to selectively bind to surface markers, enabling high sensitivity and even the capability to perform single-particle analysis. However, specific labeling may sometimes alter the native state of the EVs or introduce variability based on antibody performance. Colorimetric assays offer similar specificity and easier adaptation to PoC settings but may lack the quantitative precision and dynamic range of other methods. Electrochemical methods provide high sensitivity and rapid response times, are cost-efficient, and are well-suited for integration into portable platforms, yet their performance depends on antibody quality and immobilization characteristics. SERS-based quantification techniques, while sometimes grouped with surface marker-independent methods due to their reliance on plasmonic nanostructures, can be adapted for surface marker-specific detection. By functionalizing plasmonic substrates with specific antibodies, SERS enables both molecular fingerprinting and targeted EV recognition, achieving low detection limits and detailed biochemical information on EV cargo. However, the need for precisely engineered nanostructures and complex spectral interpretation often restricts SERS to specialized research rather than routine clinical diagnostics[66].

Microfluidic technologies offer several advantages over other analytical techniques, including minimal sample handling (i.e., no EV pre-isolation), short processing times (5-90 min), small starting volumes (hundreds of µL of biofluid), and the ability to purify less abundant EV populations. On-chip EV quantification could represent a suitable alternative to NTA, with the added benefit of data on EV sub-populations and phenotypes. While some NTA devices have a limited LOD of ~105-106 particles/mL, several microfluidic techniques demonstrate significantly lower LODs. Furthermore, microfluidic platforms can integrate sample pre-processing steps (e.g., isolation, concentration, labeling), enabling online EV quantification from whole samples. However, challenges remain in the mass-scale production of microfluidic devices, impacting reproducibility, robustness, and interuser/interlab variability. Although techniques like hot-embossing and injection molding are used for large-volume production, they are primarily suited for devices with medium or large features (above 20 µm)[67]. Smaller features limit the resolution and fidelity of mold replicas. Advances in 3D printing and roll-to-roll production are expanding the scalability of these devices, which will enhance data-backed validation of microfluidic platforms for EV quantification. Increased automation and integration with high-throughput systems would further improve reproducibility and scalability. Standardization of protocols, validation against conventional techniques, and compliance with regulatory frameworks [e.g., Food and Drug Administration (FDA) or Conformité Européenne (European Conformity, CE) marking requirements] are crucial for the widespread adoption of microfluidic devices in research and clinical applications. Additionally, the microfluidic platforms discussed in this review have been predominantly applied to serum, plasma, or whole blood. Thus, exploring their application for quantifying EVs from other bodily fluids holds potential for biomedical research, although regulatory and standardization hurdles must be addressed. It is also worth highlighting that, beyond the use of NTA as benchmarking for the development and validation of on-chip quantification, the combination of both techniques may be beneficial. Using NTA for quality sample testing, mostly when designing protocols with cell media and model cell lines, followed by on-chip quantification, may advance the field by facilitating the exploration and understanding of EVs and their clinical validation. Lastly, in light of the above-mentioned limitations, the community of EV researchers would benefit from incorporating the data generated in these studies into platforms such as EV-TRACK[68], which are specifically designed to centralize and share comprehensive data from EV-related studies. This platform includes information on the isolation and characterization techniques of EVs, as well as their origin (e.g., species or type of biofluid). For microfluidic quantification technologies, a broad classification could be established based on the main approaches suggested in this review: surface-based or surface marker-independent (nonspecific or label-free). The selection of the strategy depends on the intended application: the study of biomarkers located on the surface or contained within the internal cargo of EVs, functional studies, quantification and characterization, etc. Performance parameters, such as the LOD, capture efficiency, linear range, or reproducibility, are reporting elements that provide valuable information for researchers who may be interested in using microfluidics for their studies. Additionally, specific data on these devices could be included, such as an open repository of the microfluidic designs, fabrication methods, flow rates used, volumes sustained, operational mode, materials, or functionalization protocols. Among the most relevant aspects are the surface and/or nanoparticle functionalization methods used, whether they operate in continuous flow or perform measurements after sample processing, and whether they integrate both separation and quantification or focus solely on quantification. Addressing these challenges and incorporating growing knowledge in the EV field will facilitate the clinical and other applications of microfluidic EV analysis tools.

DECLARATIONS

Authors’ contributions

Led the conceptualization: Calero V, Honrado C, Serrano-Pertierra E

Supported the conceptualization: Ainla A, Abalde-Cela S, Macedo D, Silva H, Lage T, Diéguez L, Blanco-López MC

Investigation: Calero V, Honrado C, Ainla A, Macedo D, Silva H, Lage T, Serrano-Pertierra E

Visualization: Honrado C, Ainla A

Led the writing - riginal draft: Calero V, Honrado C, Serrano-Pertierra E

Supported the writing - original draft: Ainla A

Writing - review and editing: Calero V, Honrado C, Abalde-Cela S, Diéguez L, Blanco-López MC, Serrano-Pertierra E

Supervision: Calero V, Honrado C, Serrano-Pertierra E

Led the funding acquisition: Diéguez L, Blanco-López MC

Supported the funding acquisition: Calero V, Honrado C, Serrano-Pertierra E

Availability of data and materials

Not applicable.

Financial support and sponsorship

This work was supported by Ministerio de Ciencia e Innovación, under grant number (MCI-21-PID2020-119087RB-I00), by the European Union’s Marie Skłodowska-Curie Actions Postdoctoral Individual Fellowship programme through the grants number (101153594) (“µQuantEV”) and number (10106907) (“EVOLVE”), and by project Health from Portugal (C630926586-00465198), supported by Component C5 - Capitalisation and Business Innovation, under the Portuguese Resilience and Recovery Plan, through the NextGenerationEU Fund.

Conflicts of interest

All authors declared that there are no conflicts of interest.

Ethical approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Copyright

© The Author(s) 2025.

REFERENCES

1. Wolf P. The nature and significance of platelet products in human plasma. Br J Haematol. 1967;13:269-88.

2. Crawford N. The presence of contractile proteins in platelet microparticles isolated from human and animal platelet-free plasma. Br J Haematol. 1971;21:53-69.

3. Nunez EA, Wallis J, Gershon MD. Secretory processes in follicular cells of the bat thyroid. 3. The occurrence of extracellular vesicles and colloid droplets during arousal from hibernation. Am J Anat. 1974;141:179-201.

4. Roy S, Hochberg FH, Jones PS. Extracellular vesicles: the growth as diagnostics and therapeutics; a survey. J Extracell Vesicles. 2018;7:1438720.

5. Ramirez MI, Amorim MG, Gadelha C, et al. Technical challenges of working with extracellular vesicles. Nanoscale. 2018;10:881-906.

6. Welsh JA, Goberdhan DCI, O’Driscoll L, et al; MISEV Consortium. Minimal information for studies of extracellular vesicles (MISEV2023): from basic to advanced approaches. J Extracell Vesicles. 2024;13:e12404.

7. Ferguson S, Yang KS, Weissleder R. Single extracellular vesicle analysis for early cancer detection. Trends Mol Med. 2022;28:681-92.

8. Escola JM, Kleijmeer MJ, Stoorvogel W, Griffith JM, Yoshie O, Geuze HJ. Selective enrichment of tetraspan proteins on the internal vesicles of multivesicular endosomes and on exosomes secreted by human B-lymphocytes. J Biol Chem. 1998;273:20121-7.

9. Hallal S, Tűzesi Á, Grau GE, Buckland ME, Alexander KL. Understanding the extracellular vesicle surface for clinical molecular biology. J Extracell Vesicles. 2022;11:e12260.

10. Rupert DLM, Claudio V, Lässer C, Bally M. Methods for the physical characterization and quantification of extracellular vesicles in biological samples. Biochim Biophys Acta Gen Subj. 2017;1861:3164-79.

11. Whitesides GM. The origins and the future of microfluidics. Nature. 2006;442:368-73.

12. Convery N, Gadegaard N. 30 years of microfluidics. Micro Nano Eng. 2019;2:76-91.

13. Gholizadeh S, Shehata Draz M, Zarghooni M, et al. Microfluidic approaches for isolation, detection, and characterization of extracellular vesicles: current status and future directions. Biosens Bioelectron. 2017;91:588-605.

14. Su W, Li H, Chen W, Qin J. Microfluidic strategies for label-free exosomes isolation and analysis. TrAC Trends Anal Chem. 2019;118:686-98.

15. Chen M, Lin S, Zhou C, Cui D, Haick H, Tang N. From conventional to microfluidic: progress in extracellular vesicle separation and individual characterization. Adv Healthc Mater. 2023;12:e2202437.

16. Contreras-Naranjo JC, Wu HJ, Ugaz VM. Microfluidics for exosome isolation and analysis: enabling liquid biopsy for personalized medicine. Lab Chip. 2017;17:3558-77.

17. Ortega-sanchez FG, Teresa V, Widmann T, et al. Microfluidic systems in extracellular vesicles single analysis. A systematic review. TrAC Trends Anal Chem. 2023;159:116920.

18. Shao H, Im H, Castro CM, Breakefield X, Weissleder R, Lee H. New technologies for analysis of extracellular vesicles. Chem Rev. 2018;118:1917-50.

19. Bjørge IM, Kim SY, Mano JF, Kalionis B, Chrzanowski W. Extracellular vesicles, exosomes and shedding vesicles in regenerative medicine - a new paradigm for tissue repair. Biomater Sci. 2017;6:60-78.

20. Onukwugha NE, Kang YT, Nagrath S. Emerging micro-nanotechnologies for extracellular vesicles in immuno-oncology: from target specific isolations to immunomodulation. Lab Chip. 2022;22:3314-39.

21. Min L, Wang B, Bao H, et al. Advanced nanotechnologies for extracellular vesicle-based liquid biopsy. Adv Sci. 2021;8:e2102789.

22. Zhao Z, Fan J, Hsu YS, Lyon CJ, Ning B, Hu TY. Extracellular vesicles as cancer liquid biopsies: from discovery, validation, to clinical application. Lab Chip. 2019;19:1114-40.

23. Zhao W, Zhang L, Ye Y, et al. Microsphere mediated exosome isolation and ultra-sensitive detection on a dielectrophoresis integrated microfluidic device. Analyst. 2021;146:5962-72.

24. Yang Z, Atiyas Y, Shen H, et al. Ultrasensitive single extracellular vesicle detection using high throughput droplet digital enzyme-linked immunosorbent assay. Nano Lett. 2022;22:4315-24.

25. Lee K, Fraser K, Ghaddar B, et al. Multiplexed profiling of single extracellular vesicles. ACS Nano. 2018;12:494-503.

26. Paganini C, Hettich B, Kopp MRG, et al. Rapid characterization and quantification of extracellular vesicles by fluorescence-based microfluidic diffusion sizing. Adv Healthc Mater. 2022;11:e2100021.

27. Chen YS, Ma YD, Chen C, Shiesh SC, Lee GB. An integrated microfluidic system for on-chip enrichment and quantification of circulating extracellular vesicles from whole blood. Lab Chip. 2019;19:3305-15.

28. Chen Y, Chen C, Lai CP, Lee G. Isolation and digital counting of extracellular vesicles from blood via membrane-integrated microfluidics. Sens Actuators B Chem. 2022;358:131473.

29. Liu C, Xu X, Li B, et al. Single-exosome-counting immunoassays for cancer diagnostics. Nano Lett. 2018;18:4226-32.

30. Walker SN, Lucas K, Dewey MJ, et al. Rapid assessment of biomarkers on single extracellular vesicles using “catch and display” on ultrathin nanoporous silicon nitride membranes. Small. 2024;21:e2405505.

31. Wu Q, Wang W, Zhang C, et al. Capturing nascent extracellular vesicles by metabolic glycan labeling-assisted microfluidics. Nat Commun. 2023;14:6541.

32. Zhou S, Hu T, Zhang F, et al. Integrated microfluidic device for accurate extracellular vesicle quantification and protein markers analysis directly from human whole blood. Anal Chem. 2020;92:1574-81.

33. Ji C, Zhang M, Wang X, et al. Monolayer-fluorescence counting for ultrasensitive detection of tumour cell-derived extracellular vesicles using a step-wedge microfluidic platform. Sens Actuators B Chem. 2025;423:136786.

34. Wang Y, Gao W, Feng B, Shen H, Chen X, Yu S. Surface protein analysis of breast cancer exosomes using visualized strategy on centrifugal disk chip. Int J Biol Macromol. 2024;280:135651.

35. Vaidyanathan R, Naghibosadat M, Rauf S, et al. Detecting exosomes specifically: a multiplexed device based on alternating current electrohydrodynamic induced nanoshearing. Anal Chem. 2014;86:11125-32.

36. Stollmann A, Garcia-Guirado J, Hong JS, et al. Molecular fingerprinting of biological nanoparticles with a label-free optofluidic platform. Nat Commun. 2024;15:4109.

37. Cavallaro S, Horak J, Hååg P, et al. Label-free surface protein profiling of extracellular vesicles by an electrokinetic sensor. ACS Sens. 2019;4:1399-408.

38. Cavallaro S, Hååg P, Sahu SS, et al. Multiplexed electrokinetic sensor for detection and therapy monitoring of extracellular vesicles from liquid biopsies of non-small-cell lung cancer patients. Biosens Bioelectron. 2021;193:113568.

39. Talebian Gevari M, Sahu SS, Stridfeldt F, et al. Design and optimization of a silicon-based electrokinetic microchip for sensitive detection of small extracellular vesicles. ACS Sens. 2024;9:2935-45.

40. Kim S, Song J, Roh SM, et al. Efficient exosome separation utilizing dielectrophoretic force in conductive spiral microfluidic chips and validation via a reduced graphene oxide (rGO)-based biosensor. Sens Actuators B Chem. 2024;404:135207.

41. Wang Y, Gao W, Sun M, et al. A filter-electrochemical microfluidic chip for multiple surface protein analysis of exosomes to detect and classify breast cancer. Biosens Bioelectron. 2023;239:115590.

42. Li M, Yang P, Wu J, et al. Highly efficient and label-free sensitive detection of tumor-derived exosome with an aptasensor-based microfluidic chip. Microchem J. 2024;203:110875.

43. Qian Q, Wei Y, Xu Y, et al. Microfluidic magnetic detection system combined with a DNA framework-mediated immune-sandwich assay for rapid and sensitive detection of tumor-derived exosomes. Microsyst Nanoeng. 2023;9:139.

44. Wang J, Kao Y, Zhou Q, et al. An integrated microfluidic-SERS platform enables sensitive phenotyping of serum extracellular vesicles in early stage melanomas. Adv Funct Mater. 2022;32:2010296.

45. Zhou Q, Niu X, Zhang Z, et al. Glycan profiling in small extracellular vesicles with a SERS microfluidic biosensor identifies early malignant development in lung cancer. Adv Sci. 2024;11:e2401818.

46. Ho KHW, Lai H, Zhang R, et al. SERS-based droplet microfluidic platform for sensitive and high-throughput detection of cancer exosomes. ACS Sens. 2024;9:4860-9.

47. Hao N, Pei Z, Liu P, et al. Acoustofluidics-assisted fluorescence-SERS bimodal biosensors. Small. 2020;16:e2005179.

48. Friedrich R, Block S, Alizadehheidari M, et al. A nano flow cytometer for single lipid vesicle analysis. Lab Chip. 2017;17:830-41.

49. Gustafson KT, Huynh KT, Heineck D, et al. Automated fluorescence quantification of extracellular vesicles collected from blood plasma using dielectrophoresis. Lab Chip. 2021;21:1318-32.

50. Ware JP, Shea DK, Nicholas SL, Stimson EA, Riesterer JL, Ibsen SD. Recovery and analysis of bacterial membrane vesicle nanoparticles from human plasma using dielectrophoresis. Biosensors. 2024;14:456.

51. Hong C, Yang S, Ndukaife JC. Exosomes trapping, manipulation and size-based separation using opto-thermo-electrohydrodynamic tweezers. Nanoscale Adv. 2023;5:2973-8.

52. Kim JS, Kwon SY, Lee JY, et al. High-throughput multi-gate microfluidic resistive pulse sensing for biological nanoparticle detection. Lab Chip. 2023;23:1945-53.

53. Young TW, Kappler MP, Hockaden NM, Carpenter RL, Jacobson SC. Characterization of extracellular vesicles by resistive-pulse sensing on in-plane multipore nanofluidic devices. Anal Chem. 2023;95:16710-6.

54. Calado MRC, Lage TC, André DAM, et al. Nanofluidic resistive pulse sensing for characterization of extracellular vesicles. Lab Chip. 2024;24:4028-38.

55. Cimorelli M, Nieuwland R, Varga Z, van der Pol E. Standardized procedure to measure the size distribution of extracellular vesicles together with other particles in biofluids with microfluidic resistive pulse sensing. PLoS One. 2021;16:e0249603.

56. Jalali M, Isaac Hosseini I, AbdelFatah T, et al. Plasmonic nanobowtiefluidic device for sensitive detection of glioma extracellular vesicles by Raman spectrometry. Lab Chip. 2021;21:855-66.

57. Li D, Zou S, Huang Z, Sun C, Liu G. Isolation and quantification of L1CAM-positive extracellular vesicles on a chip as a potential biomarker for Parkinson’s disease. J Extracell Vesicles. 2024;13:e12467.

58. Ramshani Z, Zhang C, Richards K, et al. Extracellular vesicle microRNA quantification from plasma using an integrated microfluidic device. Commun Biol. 2019;2:189.

59. Sung CY, Huang CC, Chen YS, Hsu KF, Lee GB. Isolation and quantification of extracellular vesicle-encapsulated microRNA on an integrated microfluidic platform. Lab Chip. 2021;21:4660-71.

60. Cheng HL, Fu CY, Kuo WC, et al. Detecting miRNA biomarkers from extracellular vesicles for cardiovascular disease with a microfluidic system. Lab Chip. 2018;18:2917-25.

61. Zhang P, Crow J, Lella D, et al. Ultrasensitive quantification of tumor mRNAs in extracellular vesicles with an integrated microfluidic digital analysis chip. Lab Chip. 2018;18:3790-801.

62. Ko J, Wang Y, Sheng K, Weitz DA, Weissleder R. Sequencing-based protein analysis of single extracellular vesicles. ACS Nano. 2021;15:5631-8.

63. Lin H, Li B, Guo J, et al. Simultaneous detection of membrane protein and mRNA at single extracellular vesicle level by droplet microfluidics for cancer diagnosis. J Adv Res. 2024;Epub ahead of print.

64. Tong Z, Xu X, Shen C, et al. All-in-one multiple extracellular vesicle miRNA detection on a miniaturized digital microfluidic workstation. Biosens Bioelectron. 2025;270:116976.

65. Reynolds DE, Pan M, Yang J, et al. Double digital assay for single extracellular vesicle and single molecule detection. Adv Sci. 2023;10:e2303619.

66. Mahmoud AY, Teixeira A, Aranda M, et al. Will data analytics revolution finally bring SERS to the clinic? TrAC Trends Anal Chem. 2023;169:117311.

67. Scott SM, Ali Z. Fabrication methods for microfluidic devices: an overview. Micromachines. 2021;12:319.

68. Van Deun J, Mestdagh P, Agostinis P, et al; EV-TRACK Consortium. EV-TRACK: transparent reporting and centralizing knowledge in extracellular vesicle research. Nat Methods. 2017;14:228-32.

Cite This Article

Review
Open Access
Microfluidic devices for on-chip quantification of extracellular vesicles and associated biomarkers

How to Cite

Download Citation

If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. Simply select your manager software from the list below and click on download.

Export Citation File:

Type of Import

Tips on Downloading Citation

This feature enables you to download the bibliographic information (also called citation data, header data, or metadata) for the articles on our site.

Citation Manager File Format

Use the radio buttons to choose how to format the bibliographic data you're harvesting. Several citation manager formats are available, including EndNote and BibTex.

Type of Import

If you have citation management software installed on your computer your Web browser should be able to import metadata directly into your reference database.

Direct Import: When the Direct Import option is selected (the default state), a dialogue box will give you the option to Save or Open the downloaded citation data. Choosing Open will either launch your citation manager or give you a choice of applications with which to use the metadata. The Save option saves the file locally for later use.

Indirect Import: When the Indirect Import option is selected, the metadata is displayed and may be copied and pasted as needed.

About This Article

© The Author(s) 2025. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, sharing, adaptation, distribution and reproduction in any medium or format, for any purpose, even commercially, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Data & Comments

Data

Views
38
Downloads
3
Citations
0
Comments
0
0

Comments

Comments must be written in English. Spam, offensive content, impersonation, and private information will not be permitted. If any comment is reported and identified as inappropriate content by OAE staff, the comment will be removed without notice. If you have any queries or need any help, please contact us at [email protected].

0
Download PDF
Share This Article
Scan the QR code for reading!
See Updates
Contents
Figures
Related
Extracellular Vesicles and Circulating Nucleic Acids
ISSN 2767-6641 (Online)
Follow Us

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/