Perspective  |  Open Access  |  1 Aug 2023

Development of soft dry electrodes: from materials to structure design

Views: 633 |  Downloads: 91 |  Cited:   1
Soft Sci 2023;3:27.
10.20517/ss.2023.16 |  © The Author(s) 2023.
Author Information
Article Notes
Cite This Article

Abstract

Bioelectric signals reflect our daily physiological activities, which can be recorded in the form of electroencephalography, electrocardiography, electromyography, etc. The traditional Ag/AgCl wet electrode is the gold standard for clinical monitoring of bioelectrical signals at present, while complicated preparation and gel evaporation limit its long-term application. Therefore, it is meaningful to research dry electrodes without conductive paste or additional adhesives. Unfortunately, the high interface impedance between electrodes and skin is a fatal defect of dry electrodes, which leads to excessive noise levels and poor signal quality. Consequently, more efforts are required to achieve conformal contact between dry electrodes and skin to reduce the contact impedance. From this perspective, we review the recent progress in capacitive electrodes, invasive microneedle electrodes, and common-contact dry electrodes. Material selection and structural design to obtain conformal contact are highlighted. Finally, we propose the future development direction of dry electrodes.

Keywords

Dry electrode, conformal contact, electrode-skin impedance, bioelectric signal

INTRODUCTION

Extensive and complex electrical signals in organisms are the reflection of their normal physiological activities. The common bioelectric signals collected in the clinic mainly include electrocardiography (ECG), which is closely related to cardiac diseases such as arrhythmia and myocardial infarction, electroencephalography (EEG), which is closely related to brain diseases such as Parkinson's disease and stroke, and electromyography (EMG), which is closely related to muscle atrophy[1]. Dry electrodes, after years of painstaking investigation, have emerged as crucial tools for monitoring bioelectric signals[2-4]. They have been widely applied in disease prevention and diagnosis[5,6], healthcare[7-9], and human-machine interfaces[10,11].

Using dry electrodes can avoid the disadvantages of the traditional Ag/AgCl wet electrodes. For instance, the usage of wet electrodes involves complex preparation work, including hair removal and grinding the horny layer[12]. Also, evaporation of the liquid phase of wet electrodes can lead to a severe loss of signal. However, unlike wet electrodes, dry electrodes are difficult to form conformal contact with the skin due to the lack of sufficient electrolytes. This leads to high interface impedance (more than several hundreds of kΩ[13,14]), in sharp contrast with wet electrodes (a few kΩ)[15]. Besides that, improper contact will also make dry electrodes slide inevitably on the skin, causing instability in monitoring electrophysiological signals. Therefore, achieving conformal contact between dry electrodes and skin is of vital importance.

Currently, dry electrodes can be divided into the following three types according to the contact between dry electrodes and skin [Figure 1]: capacitive electrodes (non-contacted dry electrodes), invasive microneedle electrodes, and common-contact dry electrodes[16,17]. Since interface contact affects impedance and plays a decisive role in the detection performance of electrodes, this review focuses on the latest research progress in the achievement of conformal contact for three types of dry electrodes. Three types of common-contact dry electrodes, namely self-adhesive dry electrodes, ultra-thin dry electrodes, and micro-structured dry electrodes, are elaborated in terms of material selection and structural design. Then, potential challenges and future development of dry electrodes towards practical application through bioelectric signals monitoring are discussed.

Development of soft dry electrodes: from materials to structure design

Figure 1. (A-D) Schematic diagram of contact between electrodes (the wet electrode, the capacitive electrode, the invasive microneedle electrode, and the common-contact dry electrode, respectively) and skin[16]. Reprinted with permission. Copyright 2021, MDPI.

CAPACITIVE ELECTRODES

The capacitive electrode can collect signals without contacting the skin, which is similar to a capacitor and directly coupled with the skin, so it can avoid damaging the skin and measure electrophysiological signals in areas with more hair or through multiple layers of clothing[18]. However, considering the influence of coupling impedance and multi-layer insulation, conformal contact between dry electrodes and skin is not possible. Additionally, extremely high interface resistance and incomplete fixation are major obstacles for capacitive dry electrodes, leading to serious motion artifacts and noise[16]. Hence, the development of the capacitive electrode in conformal contact is actually limited, and it is not suitable for bioelectric signal monitoring, especially for dynamic monitoring.

INVASIVE MICRONEEDLE ELECTRODES

It is an important feature of the invasive microneedle electrode to use microneedles to pierce the horny layer and directly contact the growth layer of human skin in the process of monitoring electrical signals. Penetrating the horny layer formed by dead cells obtains superb conformal contact, avoiding unstable signal recording due to skin stretching or electrode movement. This method does not injure internal tissues and eliminates the high impedance of the horny layer[19,20]. Invasive microneedle electrodes can be roughly divided into dry electrodes with rigid microneedle arrays (silicon microneedle array and metal microneedle array) and dry electrodes with flexible microneedle arrays (polymer microneedle array)[17]. At present, silicon microneedles are fragile, so their development is limited. Although the metal microneedle has fine toughness, its rigid substrate is difficult to contact skin conformably. Polymer microneedles have good biocompatibility, but a few polymers with lower modulus are challenging to penetrate the horny layer. In addition, for human beings, the skin is the largest organ. Tough and soft skin has certain stretchability and elasticity[21]. As a result, owing to actualizing conformal contact between electrodes and skin, the best way is to develop a dry electrode that matches the Young's modulus of the skin. This can be achieved by using an invasive microneedle electrode with a soft substrate (lower modulus) and hard microneedle (higher modulus)[22,23]. The flexible substrate can comply with skin curvature to attain conformal contact, and the rigid microneedle can allow itself to pierce the horny layer smoothly at the same time.

Li et al. reported a dry electrode based on the polyimide microneedle array (PI-MNA) [Figure 2A]. Ti was used as an adhesive layer, and Au was used as a conductive layer, both of which were deposited onto the surface of PI-MNA by magnetron sputtering. The flexible substrate (thickness of about 50 μm) and solid microneedles (diameter less than 10 μm) of the electrode met the development of emerging invasive microneedle electrodes. Furthermore, the normalized electrode-skin contact impedance of PI-MNA electrodes had not only hit all-time lows (approximately 1/250 of that of standard electrodes), but the cost of $0.35 per electrode ensured industrial preparation of large quantities[24]. When ambient temperature or body temperature rises, sweat inevitably is produced on human skin, resulting in the accumulation of liquid at the interface of conformal contact. This not only affects signal output but also triggers skin allergies, so it is essential to ensure the permeability of dry electrodes. Hou et al. fabricated a dry electrode with microneedles at the top and Miura-ori structured substrate by polydimethylsiloxane (PDMS) molding process using a high-precision machining direct molding mold [Figure 2B]. The Miura-ori, which originated from origami, was a typical flat-folded mosaic structure. During the smooth unfolding of the origami, the parallelogram, as a component unit, remained flat and unbent. The Miura-ori structure at the bottom of the electrode not only had air circulation channels to help sweat escape but also had two-directional in-plane bendability (bending radius of 1 cm) to accommodate the curvature of skin. Compared to wet electrodes, this electrode showed an improvement in EMG signal quality (signal-to-noise ratio is 15.24 dB; conversely, the wet electrode has a signal-to-noise ratio of 7.79 dB). However, the preparation process of invasive microneedle electrodes with a Miura-ori structure was more complex. The complex structure made the homogeneity of metal plating poor[25]. Nowadays, invasive microneedle electrodes mostly adopt vertical microneedles. Interestingly, inspired by the thorn of golden margined century plant leaf, Li et al. successfully prepared the dry electrode with hook-type microneedles [Figure 2C]. The microneedle structure was fabricated by laser milling means and then embedded inside the flexible PDMS substrate. The hook-type microneedle had a smaller penetration force (0.939 N) but a larger pullout force (0.149 N), so it was laborious to break and detach[26]. Besides that, Niu et al. proposed a new stacking template technique for the first time to rapidly prepare Ag/AgCl thermoplastic polyurethane (TPU) electrodes [Figure 2D]. They interleaved and fixed the bent pins, which eventually became a multi-stacked stencil. TPU was poured into the templates for curing, and silver conductive layers were deposited by chemical silver plating. The electrode had a structure similar to the inverted thorn of Xanthium and an excellent dynamic friction coefficient (about 38.8% higher than that of a flat plate structure). Accordingly, this electrode showed a high signal-to-noise ratio in dynamic signal acquisition (32.27, 30.13, and 27.77 dB when the arm was swung at 30, 60, and 90; conversely, the signal-to-noise ratio of the wet electrode was 31.89, 26.73, and 21.92 dB)[27]. In fact, the hooked curved electrode does provide better conformal contact and lower contact impedance compared to the upright microneedle electrode. However, the curved needle that tightly binds to the skin is difficult to pull out after testing and is prone to cause greater skin damage, which limits its practical application.

Development of soft dry electrodes: from materials to structure design

Figure 2. Invasive microneedle electrodes. (A) SEM image of the polyimide microneedle array electrode and its sample diagram[24]. Reprinted with permission. Copyright 2022, Springer Nature. (B) SEM image of the Miura-ori structured electrode (illustration inside is an enlarged SEM image of a single microneedle) and its sample diagram[25]. Reprinted with permission. Copyright 2021, Springer Nature. (C) Size and shape drawing of the hook electrode and its sample diagram[26]. Reprinted with permission. Copyright 2020, Elsevier. (D) SEM image of the barbed electrode and its physical picture[27]. Reprinted with permission. Copyright 2022, American Chemical Society. SEM: Scanning Electron Microscope.

For invasive microneedle electrodes, softness is an important prerequisite for their further development in the field of conformal contact. Flexible microneedle electrodes can adapt to skin profiles to obtain lower signal-to-noise ratios in dynamic measurements. Nevertheless, as the manufacturing engineering of invasive microneedle electrodes becomes increasingly mature, researchers gradually discover that microneedles of any material all have a sharp structure, bringing discomfort and infection. This is also the challenge that limits the practical application of invasive microneedle electrodes.

COMMON-CONTACT DRY ELECTRODES

The common-contact dry electrode means that it can directly contact skin but does not pierce the horny layer. Compared with the invasive microneedle electrode, it cannot completely adapt to the rough skin surface, thus resulting in high and unstable interface impedance[17]. In order to gain conformal contact, scientists improved traditional dry electrodes and invented the following types of the common-contact dry electrode, including self-adhesive dry electrodes, ultra-thin dry electrodes, and micro-structured dry electrodes.

Self-adhesive dry electrodes

To achieve conformal contact between dry electrodes and skin, it is an effective approach to endow dry electrodes with self-adhesive characteristics. Poly (3, 4-ethylenedioxythiophene): poly (styrene sulfonate) (PEDOT:PSS) is a polymer material with the advantages of electrical conductivity, transparency, and environmental stability, which is often used in the field of flexible electronics. From Ouyang’s work, a stretchable and self-adhesive dry electrode (PWS) was reported via solution processing of the biocompatible mixture of PEDOT:PSS, waterborne polyurethane (WPU), and D-sorbitol [Figure 3A]. Intrinsically conductive PEDOT:PSS did not have good elasticity and adhesion. O-H, N-H, and C=O groups derived from WPU and D-sorbitol were able to form strong physical adsorption (hydrogen bonding) with keratin and lipids on the skin. Naturally, the PWS electrode exhibited lower contact impedance (82  KΩ cm2 at 10  Hz) than the Ag/AgCl electrode (148  KΩ cm2 at 10  Hz), which could accurately record high-quality ECG signals (peak-to-peak voltage of 1.84 mV for QRS complex)[28]. Although the self-adhesive dry electrode is capable of detecting high-quality signals in motion, its adhesion is greatly deteriorated by sweat. Consequently, future research should focus on dry electrodes with high self-adhesive properties for wet skin. Cao et al. proposed a self-adhesive and highly sweat-resistant dry electrode (PPT) by drop-casting aqueous solution. The electrode was the blends of PEDOT:PSS, polyvinyl alcohol (PVA), tannic acid (TA), and ethylene glycol (EG) [Figure 3B]. Fundamentally, the increase in conductivity (122 Scm-1) attributed to EG was able to induce secondary doping of PEDOT:PSS. As a soft polymer, PVA could disperse energy and improve the breaking elongation (54%) of the electrode. The hydroxyl group and hydrophobic benzene ring of TA could strongly increase the adhesion of the PPT electrode in dry and clammy environments (0.32 and 0.28 Ncm−1, respectively)[29]. Similarly, Park et al. chose PEDOT:PSS as the conductive material to make a degradable and environment-friendly dry electrode (PDC) by blending the component solutions [Figure 3C]. It was worth mentioning that they added dopamine and hydroxyethyl cellulose (HEC) into the PDC film. The amine group of dopamine formed a hydrogen bond and electrostatic interaction with the hydroxyl group on HEC, giving the PDC electrode excellent tensile properties (only 8% of resistance change under 110 of fully flat-bended cycles). More importantly, HEC could be dissolved in water, so this electrode was able to be degraded under 60 °C distilled water immersion or washing[30]. In fact, the degradable property enables the post-processing of electrodes and increases the sustainability of the product, which has profound application value.

Development of soft dry electrodes: from materials to structure design

Figure 3. Self-adhesive dry electrodes. (A) Preparation flow chart of the electrode based on PEDOT:PSS (the illustration in the lower right corner is its physical picture)[28]. Reprinted with permission. Copyright 2020, Springer Nature. (B) Preparation diagram of the highly sweat-resistant dry electrode and photos of its stretch/compression at the wrist[29]. Reprinted with permission. Copyright 2022, American Chemical Society. (C) Formation diagram of the environment-friendly dry electrode and its photos showing various characteristics[30]. Reprinted with permission. Copyright 2022, Wiley-VCH. (D) Explanatory diagram of the interface adhesion mechanism of the electrodes based on LM-doped elastomers and its SEM image (the illustration in the upper right corner is a sample picture)[37]. Reprinted with permission. Copyright 2022, Wiley-VCH. (E) Pictures of PPy and SF before and after interfacial polymerization and a sample picture of the electrode based on this interface interlocking structure[40]. Reprinted with permission. Copyright 2020, American Chemical Society. (F) Structure diagram of the electrode with a three-layer structure and its contact pictures with the skin under different conditions[41]. Reprinted with permission. Copyright 2022, Elsevier. LM: Liquid metal; PEDOT:PSS: poly (3, 4-ethylenedioxythiophene): poly (styrene sulfonate); PPy: polypyrrole; SF: silk fibroins.

Apart from PEDOT:PSS[31-33], elastomers that inherently possess viscous properties are often used as substrate materials for self-adhesive dry electrodes. However, poor conductivity is a fatal defect of elastomers. Therefore, enhancing the electrical conductivity of elastomers is the key to their wide application in self-adhesive dry electrodes. Generally speaking, the common method to elevate the conductivity of elastomers is to add conductive fillers, such as metal nanoparticles, metal nanowires, and carbon nanotubes. Nevertheless, the traditional conductive fillers will not maintain a conductive network under strain, resulting in a rapid decline in conductivity. In recent years, liquid metals (LMs) with excellent conductivity and ductility have gradually come into people's view. Moreover, LMs with tensile properties matching elastomers enable them to overcome the drawback of conventional fillers[34-36]. Pei et al. applied LMs as fillers to endow viscoelastic elastomer with high tensile conductivity [Figure 3D]. EGaIn droplets were uniformly dispersed into the uncured elastomer precursor composed of a catechol group and thiooctyl monomer. On account of the existence of the LM, hydrophobic alkyl chain, dynamic disulfide bond, catechol group, and carboxyl group in the system, this electrode had high conductivity (1.3×104 Sm-1), outstanding reversible wet bonding strength (670 kPa), and fine self-healing function under high temperature/light (90% healing efficiency after 4 hours of healing at 60 oC)[37]. Practically, the self-healing property gives elastomers the ability for long-term monitoring, which further expands the application of dry electrodes.

Dry electrodes are applied to the delicate skin of the human body. Once electrodes are toxic or irritating, they can lead to skin damage and cause discomfort to the user. Therefore, biocompatibility is an important criterion for measuring dry electrodes. Some natural adhesive materials, such as silk fibroins (SF), have attracted people's attention because of their exceptional biocompatibility and superior conformal adhesion[38,39]. Yang et al. obtained the structural interlock of Polypyrrole (PPy) and SF by means of interfacial polymerization and fabricated a self-adhesive dry electrode with water-activated conformal properties [Figure 3E]. At higher relative humidity (94.6%), SF with Young's modulus (2.1 MPa) close to that of skin could boost interfacial adhesion (40 J/m2). Meanwhile, the chain structure of PPy and SF promoted the tensile stability of this electrode (R/R0 at 100 cycles of stretch was approximately 1). These features ensured that this electrode was capable of collecting stable and reliable ECG signals in the environment of sweating and exercise (sensitivity of 0.5; conversely, the sensitivity of the wet electrode was 0.31)[40]. Additionally, a self-adhesive dry electrode (PPy@AM-SF/CNC) with low impedance was prepared by Meng et al. and was able to easily detach from the skin [Figure 3F]. The group used acid-modified silk/cellulose nanocrystal film as the substrate, coated PPy as the conductive layer, and finally spread calcium-modified SF as the adhesive layer. Among these three layers, the SF layer was modified with Ca to facilitate the adhesion of the PPy@AM-SF/CNC electrode (12.85 N/m on pigskin) because Ca ions could chelate with metals and trap water. Notably, when the electrode was immersed in water, the cohesive force decreased due to the obvious increase of water content (peel force dropped sharply from 13.5 N/m to 9.22 N/m in 10 s), so this effectively achieved a non-destructive separation from skin[41].

Adhesion of self-adhesive dry electrodes is profitable for their conformal contact, thereby reducing skin contact impedance and noise. However, the adhesion of electrodes on wet skin is not ideal, so the development of self-adhesive dry electrodes that can be applied to both wet and dry skin is a future endeavor. From the perspective of materials, PEDOT:PSS and elastomers are common substrate materials for the fabrication of stretchable self-adhesive electrodes. Although their intrinsic properties are limited, this is compensated by the doping of certain substances (TA, LMs, etc.). Additionally, with the increasing research on the practical application of dry electrodes, people are paying more attention to the biocompatibility of electrodes. Some biocompatible viscous substances (SF) can help us to obtain self-adhesive dry electrodes with little skin damage. However, poor stretchability and low conductivity of viscous substances limit them to further development. Hence, the modification of natural viscous substances is an important trend for self-adhesive dry electrodes.

Ultra-thin dry electrodes

It is believed that thick electrodes are hard to adhere tightly to the skin[42]. This leads to a decrease in the contact area at the electrode-skin interface, affecting the quality of signals. Specifically speaking, for the same material, thick electrodes are usually more difficult to bend than ultra-thin electrodes. Therefore, it cannot establish a wonderfully conformal contact with skin, which is more prone to generate noise and artifacts. On the contrary, the ultra-thin electrode with better conformability is capable of filling the wrinkles and depressions of the skin, thus maintaining a larger contact area and obtaining better signal quality[43]. In brief, reducing the thickness of electrodes is an uncomplicated and valid tactic of conformal contact. However, the imbalance between the mechanical and electrical properties of ultra-thin electrodes is a common conundrum in research. From this, Tang et al. utilized solution processing of PEDOT:PSS, polysorbate, and glycerol to fabricate a delamination-resistant imperceptible dry electrode [Figure 4A]. The polar group in glycerol could lead to the phase separation of PEDOT and PSS; that is to say, the partial aggregation of PEDOT formed a conductive network, and PSS became a dispersed soft matrix. As a functional bridging unit of PEDOT and PSS, the biocompatible polysorbate could dissipate the tensile force through its hydrophilic/hydrophobic interaction, which endowed the electrode with high fracture strain (elongation at break was higher than 90%, and the corresponding resistance increased only 1.7 times). More importantly, the thickness of several microns enhanced the compliance of the electrode to the skin, obtaining a good signal-to-noise ratio for EMG (35.23 dB)[44]. Jiang et al. evaporated the gold film onto the PDMS film and proposed a PDMS-gold conductor [Figure 4B] with a thickness of only 1.3 microns. During the preparation process, the thick PDMS film expanded, and the thin gold film emerged with controlled micro-cracks. This micro-cracked structure maintained steady conductivity of the electrode during stretching (resistance only increases by 1.7% at 0% strain after 5000 cycles at 100% strain). Furthermore, the electrode could record ECG signals with a high signal-to-noise ratio (37.5 dB)[45].

Development of soft dry electrodes: from materials to structure design

Figure 4. Ultra-thin dry electrodes. (A) The physical picture of the delamination-resistant imperceptible dry electrode and its cross-section optical microscope image after contacting pig skin[44]. Reprinted with permission. Copyright 2021, American Chemical Society. (B) Structure diagram of PDMS-gold conductor and its sample photo[45]. Reprinted with permission. Copyright 2022, Springer Nature. (C) Schematic diagram of the single-layer porous membrane electrode (the illustration in the upper right corner is its real photo)[48]. Reprinted with permission. Copyright 2022, American Chemical Society. (D) Description of the principle of the porous ultra-thin electrode (the illustration in the upper right corner is its physical picture)[49]. Reprinted with permission. Copyright 2022, Springer Nature. (E) Schematic diagram of hydrogen bond in the ultrathin MXene-based electrode and SEM diagram based on the pore structure of grass-derived cellulose[51]. Reprinted with permission. Copyright 2022, American Chemical Society. (F) Physical image and SEM image of the breathable electrode based on the electrospinning method[54]. Reprinted with permission. Copyright 2021, American Chemical Society. (G) Process flow diagram of the ultra-thin (10 μm) fabric electrode and its SEM diagram[55]. Reprinted with permission. Copyright 2022, Elsevier. PDMS: polydimethylsiloxane.

Because the ultra-thin electrode has excellent bonding ability with skin, bacteria are easy to breed, and signal quality is poor in sweaty and humid environments. More attention should be paid to the permeability of ultra-thin electrodes. Porosity is a means to elevate air permeability[46,47]. Tian et al. reported a single-layer porous membrane electrode [Figure 4C] using the breath figure method (BFM). N2 was blown onto the surface of the chloroform solution of styrene-ethylene-butylene-styrene (SEBS)/adhesive and amphiphilic polymer (AAP). The chloroform rapidly evaporated, resulting in a porous film structure. Then, gold was magnetically sputtered on the surface of the porous film. The pore size and wall thickness were less than 10 μm, and it was well matched with the pores of human skin, making it comparable to the ultrahigh permeability of naked skin (Approximately 25.3 gm−2d−1). Moreover, this electrode was clearly able to observe the ECG waveform and obtain a heart rate of approximately 62 bpm[48]. Xie et al. fabricated a porous ultra-thin (200 nm) electrode [Figure 4D] using the Marangoni effect and magnetron sputtering gold. Compared to water, the surface tension of SEBS solution was lower, so an ultra-thin layer of SEBS solution would appear on the water surface. The gradual evaporation of toluene increased the concentration of SEBS, resulting in an increase in surface tension and a decrease in the viscosity of SEBS. When the surface tension of the SEBS solution decreased, material transport would occur, inducing the formation of membrane pores. At the same time, the increase in the viscosity of the SEBS solution restricted the flow and the increase in pores. When the toluene was completely evaporated, an ultra-thin porous film (25.3 gm−2h−1) was accomplished. Finally, sputtering gold endowed this electrode with conductivity. The signal-to-noise ratio of this electrode during sweating was close to that of the wet electrode (about 13 dB)[49]. Additionally, Yu et al. reported a highly breathable and ultrathin MXene-based electrode (MBE) [Figure 4E]. Transition metal carbides or nitrides, known as MXene[50], are two-dimensional materials with superb conductivity and biocompatibility. Inspired by the natural porous structure of plants, they chemically etched Pennisetum to prepare a flexible porous cellulose skeleton and then impregnated the MXene sheet. Principally, the functional groups of -O, -OH, and -F in MXene could form hydrogen bonds with -COOH and -OH in cellulose. This hydrogen-bonded MXene nanosheet made MBE breathable and excellent conductive (10 Ω/sq). Surprisingly, MBE had wonderful joule heating ability, which could be used as a combination of electrical stimulation and electrotherapy functions in muscle therapy[51].

The nanofiber network structure is another way to increase air permeability. As is well-known, electrospinning is popularly used to prepare porous nanofiber network substrate[52,53]. Jeong et al. proposed a breathable electrode [Figure 4F] that sprayed PEDOT:PSS on TPU elastomer nanofiber produced by electrospinning. This structure with nano-network was the source of outstanding breathability (12.0 gm-2h-1), expanding the application prospect of ultra-thin electrodes[54]. Yan et al. fabricated an ultrathin (10 μm) fabric electrode (PAS) [Figure 4G]. They covered Ag NWs (as a conductive layer) on the surface of hydrophobic polytetrafluoroethylene (PTFE) non-woven fabric (as an external self-cleaning layer) by suction filtration operation. Then, viscous SFs (Silk NFs) (as a skin contact layer) were electrospun on the conductive layer. This PAS electrode had brilliant permeability (2553 gm-2d-1), surface hydrophobicity (contact angle 142.3°), and outstanding conductivity (3.58 Ω/sq). These brilliant characteristics made the electrode have low noise and high signal-to-noise ratios (22.59 dB). However, since Ag NWs were bonded to PTFE using physical suction filtration, the cyclic stretching conductivity of this electrode was a weak point for high-quality monitoring of long-term motion[55].

The thickness of ultra-thin electrodes is usually at the nanometer or micrometer level, which makes them better conform to the curves of skin, less prone to falling off, and less likely to create gaps at the interface. Nevertheless, it is a challenging task to balance the mechanical and electrical properties of ultra-thin electrodes while controlling the low thickness. In addition, how to enhance the permeability of ultra-thin electrodes to avoid skin inflammation is also worthy of in-depth consideration. In fact, porous membranes and fiber network structures are two means to ensure the air permeability of electrodes.

Micro-structured dry electrodes

The micro-structured electrode refers to the design of a special microstructure on the surface of electrodes so that the electrode itself can pass through the hair but not penetrate the skin. The microstructure ensures a larger contact area, lower contact impedance, and better conformal contact[56]. At present, some microstructures have been prepared extensively by researchers, such as claw-like, spring-like, micropillar-like, and bionic microstructures.

Xing et al. reported a tentacle electrode whose shape was made by molding [Figure 5A]. They chose TPU, a polymeric elastomer, as the primary manufacturing material. The electrode was designed as a claw-like structure with a diameter of 14 mm and had a hemispherical tip featuring eight small fingers of 6 mm in length and 2 mm in diameter. This structure and elasticity could help fingers pass through hair and touch skin to record EEG signals with high signal-to-noise ratios (12.83 ± 2.85 dB; conversely, a signal-to-noise ratio of 14.1 ± 3.24 dB for the wet electrode)[57]. Liu et al. developed a novel dry electrode consisting of an upper printed circuit board (PCB), a lower PCB, a thread holder, and spring probes [Figure 5B]. The buffering effects provided by spring contact probes and the flexible cap enabled dry electrodes to attach well to the scalp without pain. Simultaneously, the EEG signal of this electrode was not significantly different from that of the wet electrode (relative error between 0.8179 ~ 0.9677), which could be easily and quickly used in clinical applications[58]. Using anodic oxidation (AAO) as a template, Niu et al. reported a micro-nanopillar array-structured dry electrode [Figure 5C]. They used in situ polymerization method to polymerize PANI nanopillars on the pore wall of AAO and spin-coated TPU/N,N-dimethylformamide (DMF) solution on AAO/PANI. After curing, the AAO was removed to obtain the PANI/TPU dry electrode. Compared with the planar wet electrode structure (2.257 MΩ at 0.1 Hz; 13.54 ± 7.86 dB), the electrode with the micro-nanopillar array structure could increase the contact area, thus showing a lower contact impedance (375.5 kΩ at 0.1 Hz) with a higher ECG signal-to-noise ratio (21.33 ± 5.48 dB) in a dynamic environment as well[59].

Development of soft dry electrodes: from materials to structure design

Figure 5. Surface microstructure electrodes. (A) Image of the caw-like dry electrode[57]. Reprinted with permission. Copyright 2018, Springer Nature. (B) Schematic diagram and a sample picture of the spring-like dry electrode[58]. Reprinted with permission. Copyright 2019, Elsevier. (C) SEM image of the micronanopillar electrode and its physical photo[59]. Reprinted with permission. Copyright 2022, American Chemical Society. (D) SEM image of the grasshopper-inspired bionic dry electrode with a scale of 200 µm (200 × ) and 20 µm (2000 × )[62]. Reprinted with permission. Copyright 2018, Wiley-VCH. (E) Schematic diagram of the water-drainable dry electrode inspired by frog and octopus[63]. Reprinted with permission. Copyright 2019, Wiley-VCH. (F) SEM image of the microstructure electrode inspired by the forelegs of male diving beetles and its physical photo[64]. Reprinted with permission. Copyright 2021, Wiley-VCH. (G) SEM images from different angles of the Janus Au NWs/NC electrode inspired by the conical beak of shorebirds (the nitrocellulose side, the cross-section, and the Au NWs side from left to right, scale bars: 100 µm) and its physical picture[65]. Reprinted with permission. Copyright 2022, Wiley-VCH.

The bionic microstructure electrode fabricated by imitating the unique physical structure of organisms has several distinctive properties, such as adhesion[60] and the directional transport of water[61]. Inspired by grasshopper feet, Stauffer et al. reported a micro-structured dry electrode using the template method [Figure 5D]. This electrode was a combination of a soft porous electrode at the bottom (with an elastic modulus of ≈ 1 MPa similar to that of human skin), a macroscopic pillar electrode in the middle, and a micro-structured electrode at the top. This dry electrode could pass through thick hair and make conformal contact with skin. Furthermore, the skin contact impedance was comparable to that of clinical standard electrodes (approximately 50 kΩ cm2 at 10 Hz)[62]. However, a challenge remains in terms of the adhesion of dry electrodes in human sweating environments. To solve this problem, Kim et al. reported a water-drainable dry electrode with a hexagonal micro-pattern inspired by the hexagonal micro-pattern of a tree frog toe pad and convex cup in an octopus sucker [Figure 5E]. They first prepared silicon templates using photolithography and reactive ion etching techniques. PDMS was fabricated into microstructures using simple micro-contact printing methods and then was sprayed with reduced graphene oxide (1.2 µm in thickness). The hexagonal micro-patterned channel facilitated sweat elimination, and convex cups on its upper surface improved skin adhesion in sweaty environments (2.36 Ncm-2 in dry conditions; 1.92 Ncm-2 in sweaty conditions) so that the electrode avoided detachment from the skin in both static and dynamic environments[63]. The surface of the forelegs of male diving beetles has a micro-cavity sucker structure and micro-wrinkles under the micro-cavity. The existence of micro-cups and micro-wrinkles makes the forelegs of beetles possess strong suction, liquid drainage, and high tensile strength. Inspired by this, Min et al. fabricated a flexible electrode with isotropic folded suction cups by compounding multi-walled carbon nanotubes (MWCNT) with PDMS via a processable selective transfer technique [Figure 5F]. Specifically, MWCNT suspensions were sprayed on plasma-treated and pre-strained PDMS. After tension release, the sprayed MWCNT wrinkled PDMS. This electrode had excellent adhesion (2.2 Ncm-2 on dry pig skin; 1.5 Ncm-2 on sweaty pig skin), superb electrical properties (approximately 87 Sm-1; conductivity change of only 0.1 even at 30% tensile strain), and good moisture resistance (contact angle < 141.9°)[64]. Inspired by the conical beak of shorebirds, a sweat-proof Janus Au NWs/nitrocellulose (NC) electrode with a conical microporous structure was reported by Li et al. [Figure 5G]. Because the semi-cured polyurethane (PU) was highly adhesive, they prepared Au NWs on one side of the PU and lightly rolled the NC film onto the other side of the PU. The conical hole fabricated by the laser perforation method could spontaneously drive sweat from the electrode layer (the Au NWs side) to the sweat adsorption layer (the NC side), thus maintaining high interfacial adhesion (approximately 1.7 MPa) and low interfacial impedance (7 kΩ at 100 Hz; conversely, the impedance of bare Au NWs without microstructure is 10 kΩ) in the sweaty state[65].

With the development of micro and nanotechnology and the study of organisms in nature, people gradually focus on dry electrodes with surface microstructure. The microstructure generally has textural structure rules and certain periodicity. In addition, the microstructure can be arranged in different sizes and combinations to make electrodes exhibit different functional properties, such as adhesion, friction, and hydrophobicity. The current investigation of micro-structured dry electrodes involves multidisciplinary and has become a research hotspot. However, the development of micro-structured dry electrodes is still limited by processing technology. Thus, future research focus can be on the design and manufacturing technology of micro and nanostructures.

As shown in Table 1, among the three types of dry electrodes discussed, capacitive electrodes are able to avoid damage to the skin because they do not touch the skin. They can be applied to hairy areas and pass through multiple layers of clothing. However, they have the drawbacks of poor conformal contact with skin, the highest contact impedance, and lower quality in dynamic monitoring signals. Accordingly, capacitive electrodes are of little interest for research in the field of conformal contact. Invasive microneedle electrodes use microneedles to penetrate the horny layer, so they have better conformal contact with skin, minimal electrode-skin interface impedance, and superior signal-to-noise ratios. However, the discomfort and inflammation caused by the microneedle pose a threat to human health and limit the practical application of electrodes. Common-contact dry electrodes are characterized by direct contact with skin but do not penetrate the horny layer. This non-invasive nature determines relatively little skin damage and contact impedance, so the research prospects of common-contact contact dry electrodes are broad. However, complete conformal contact with skin is difficult, except with the help of certain properties (self-adhesive or ultra-thin) or special structures (microstructures). For self-adhesive dry electrodes, the humid environment limits their adhesion, which is not suitable for long-term monitoring of electrophysiological signals or detection under heavy sweating scenarios. Likewise, the weak breathability of ultra-thin dry electrodes easily causes skin inflammation and discomfort to users, making it difficult to detect signals on human skin for a long time. Nevertheless, there is a gap between the micro-structured dry electrode and the skin to remove gas and sweat generated by skin metabolism. Thus, inflammation is avoided, and long-term signal detection is facilitated. Actually, the signal-to-noise ratio of micro-structured dry electrodes is slightly lower than that of self-adhesive dry electrodes and ultra-thin dry electrodes, but reasonable design of microstructures can improve the contact situation to achieve the same level of signal-to-noise ratio. In addition, various organisms and phenomena in nature are our inspiration for designing microstructures. Bionic structures have more ingenious and interesting properties. Therefore, micro-structured dry electrodes have great research value and great prospects.

Table 1

Advantages and disadvantages of capacitive electrodes, invasive microneedle electrodes, and common-contact dry electrodes

TypeAdvantagesDisadvantages
Capacitive electrodesNo skin traumaSuitable for hairy and multi-layer clothing isolationNon-conformal skin contact and high contact impedance
Invasive microneedle electrodesConformal contact and low contact resistanceSkin damage and user discomfort
Common-contact dry electrodesLight skin damageIncomplete conformal contact

CONCLUSION AND OUTLOOK

Recent advances of dry electrodes are reviewed, focusing on their conformal interfacial contact. Capacitive electrodes have inferior interfacial contact and high impedance owing to the presence of a dielectric layer. Invasive microneedle electrodes have microneedles inserted into the horny layer of skin, providing excellent conformal contact but easily causing discomfort. Common-contact dry electrodes are preferred by scientists for their ability to realize conformal contact with the skin while being non-invasive. Moreover, three types of common-contact dry electrodes, namely self-adhesive dry electrodes, ultra-thin dry electrodes, and micro-structured dry electrodes, are mainly discussed. Simultaneously, their material selection and structural design to obtain conformal contact are elaborated. In addition, looking back at the development in recent years, it is not difficult to find trends and challenges of dry electrodes in the future.

During bioelectric signal measurement, the movement of human skin causes the electrode to slip on the skin and thus causes signal noise, which is the source of motion artifacts. Conformal contact with skin can reduce interface impedance and enhance signal quality. Therefore, dry electrodes in conformal contact with the skin are the best way to decrease motion artifacts. In general, stretchable, flexible, self-adhesive, ultra-thin, and micro-structured properties can be endowed to the dry electrode to achieve conformal contact with skin.

Superb stretchability and flexibility are basic ideal features for dry electrodes. Specifically, electrodes better adapt to the deformation environment with the aid of high stretchability, while high flexibility helps electrodes better comply with skin. Intrinsically stretchable materials and structural design are two strategies to achieve the stretchability of electrodes. Intrinsically stretchable materials mainly include elastomers, gels, and semiconductor polymers. Conductive materials, such as metals, are usually used as fillers to improve their conductivity. For some non-stretchable rigid materials, geometrical structures are designed to give them stretchability. For example, micro-cracks, serpentines, out-of-plane waves, open-mesh, and kirigami all exhibit outstanding anti-deformation capability[66]. Since dry electrodes work on soft skin, the electrodes must be flexible for conformal contact and wearing comfort. Considering the correlation between the Young's modulus and flexibility, the Young's modulus of electrodes is supposed to be reduced, which can smoothly conform to the skin and improve interface contact.

An effective means of maintaining conformal contact between electrodes and skin is interfacial adhesion. In the past, additional adhesives (conductive gels) were used to attain interfacial adhesion. For easy use of electrodes, scientists turn their focus to self-adhesive electrodes. However, self-adhesive organic dry electrodes always encounter the hassle of viscosity reduction in a humid environment. As a consequence, effective adhesion of self-adhesive dry electrodes on dry and wet skin is crucial for practical applications. Additionally, although strong interfacial adhesion facilitates conformal contact, electrodes, after use, are difficult to peel off from the skin and may even cause damage. Therefore, convenient detachment from the skin is also the way forward for dry electrodes.

The thickness of the electrode is closely related to the bending stiffness. Compared to thick electrodes, ultra-thin electrodes have less bending stiffness and stronger van der Waals forces at the interface, resulting in more conformal contact with skin. However, due to sweat and gas discharged from human skin, skin impedance will increase, and bacteria are more likely to breed, so we must leave no stone unturned to enhance the permeability of electrodes. Intrinsically breathable materials are poorly permeable, while the structural design can efficiently endow electrodes with excellent breathability. Examples include porous membranes prepared by blown bubbles, phase separation, and sacrificial sugar templates or fiber network membranes prepared by electrostatic spinning. In addition, ultrathin electrodes often face the problem of imbalance between mechanical and electrical conductivity, so a good balance between them is what researchers have been pursuing.

The ingenious microstructure allows electrodes to pass through hair and directly touch the skin, thus making better conformal contact and reducing contact resistance. When researchers design microstructures, inspiration should not be limited to daily life. The creatures and phenomena in nature may be helpful to us. After all, all living creatures change themselves to adapt to nature through cruel natural selection. Applying their physical characteristics and habits to electrodes can yield unexpected benefits.

Besides the above characteristics, the increasing demand for electrodes with long-term applications has made the study of self-healing and biocompatibility increasingly popular. Traditional conductive active materials (metal/conductive polymers) are prone to irreversible damage and fracture under external mechanical manipulation, so exploiting electrodes with rapid self-healing properties can effectively alleviate this problem, such as living assembled materials that take advantage of the healing properties of bacterial cell growth[67]. Biocompatibility is a constant theme in biomaterials research. Since dry electrodes are applied directly to the surface of human skin, biocompatibility is critical. Biocompatibility is a property of living body tissues that determines their reaction to a product. Better biocompatibility means that the product is less harmful to living organisms. During the design of dry electrodes, we need to consider whether the product triggers skin allergy or inflammation and whether it can remain low or even non-toxic in the long term. To improve the biocompatibility of dry electrodes, researchers should avoid using irritating and toxic substances as much as possible. Notably, the use of natural materials, such as silk protein SF and dopamine, can greatly enhance biocompatibility. Moreover, in response to the national double-carbon policy, the investigation of the degradable electrode, a kind of green product, should not be ignored.

Although significant progress has been made in the study of various aspects of performance just mentioned, the preparation of practical electrodes with excellent comprehensive performance is still a challenge. Consequently, it is imperative to achieve a perfect balance of all properties (stretchability, conductivity, flexibility, etc.).

It is worth mentioning that dry electrodes have not been actually on the market, and wet electrodes still occupy the main position. Nowadays, dry electrodes with conformal contact studied in the laboratory can reach the same level as wet electrodes in monitoring signals (signal-to-noise ratio). However, to truly become a viable commercial product, dry electrodes not only have excellent performance and mature process routes but also need to have low cost, good user experience, and long-term durability. In addition, the electrophysiological signals of the human body are unstable and complex. There is mutual crosstalk between various signals, such as mutual interference between myoelectricity and cardiac electricity. Therefore, how electrodes take specific signals while circumventing other signals needs to be further investigated by scientists. Unfortunately, we still have not found a perfect solution to coordinate all aspects of electrode performance. Can we realize large-scale commercial production in the future? This is an interesting topic for the practical application of dry electrodes. The idea that dry electrodes obtain mass commercial production merits careful consideration.

A product fulfills its specific practical value when it is applied. Designers must fully consider specific application scenarios so that they can create commercial products with maximum functionality to meet specific needs. Dry electrodes monitor electrophysiological signals to achieve health monitoring, making them highly promising in the biomedical field. For example, they are used in applications such as ECG scanning[68], respiratory monitoring[69], and pulse monitoring[70]. In the digital era, the rapid rise of artificial intelligence and the Internet of Things has led to a growing desire for human-computer interaction. The electrophysiological signals of the human body monitored by dry electrodes can be transformed into specific signals for human-computer interaction through a series of processing. Examples include hand gesture recognition[71], lip reading[72], and prosthetic control[73]. Smart wearable devices are the ideal future model for all types of electronics. Dry electrodes integrated into wearable devices capture information about the human body and do not interfere with the activities of users. Meanwhile, such smart wearable devices can transmit data with smartphones for easy monitoring[74-76]. Exercise is a normal part of life nowadays. Using dry electrodes for intelligent motion monitoring not only helps us to understand the state of the body during exercise but also provides scientific exercise guidance[77-79]. Significantly, a thorough understanding of the application scenarios of dry electrodes can lead researchers to carry out targeted designs. Furthermore, with the development of system integration and machine intelligence, the integration of dry electrodes into systems such as wearable devices or sensors[80] will be a very promising direction.

DECLARATIONS

Acknowledgments

The authors acknowledge the funding support from the National Natural Science Foundation of China (Grants No. 52173237), the Fundamental Research Funds for the Central Universities (Grants No. HIT.OCEF.2022018), the Interdisciplinary Research Foundation of HIT (IR2021207), Project of National Center for International Research on Intelligent Nano-Materials and Detection Technology in Environmental Protection, Soochow University (No. SDGH2105), and the Open Project Program (PEBM202107) of Key Laboratory for Photonic and Electric Bandgap Materials, Ministry of Education.

Authors’ contributions

Wrote the original draft: Liu H, Chen X, Liu Y, Liang C

Supervised, reviewed, and revised the manuscript: Wang Z, Zhu M, Qi D

Availability of data and materials

Not applicable.

Financial support and sponsorship

This work was supported by the National Natural Science Foundation of China (Grants No. 52173237), the Fundamental Research Funds for the Central Universities (Grants No. HIT.OCEF.2022018), the Interdisciplinary Research Foundation of HIT (IR2021207), the Project of National Center for International Research on Intelligent Nano-Materials and Detection Technology in Environmental Protection, Soochow University (No. SDGH2105) and the Open Project Program (PEBM202107) of Key Laboratory for Photonic and Electric Bandgap Materials, Ministry of Education.

Conflicts of interest

All authors declared that there are no conflicts of interest.

Ethical approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Copyright

© The Author(s) 2023.

REFERENCES

1. Niu X, Gao X, Liu Y, Liu H. Surface bioelectric dry electrodes: a review. Measurement 2021;183:109774.

2. Zhu M, Wang H, Li S, et al. Flexible electrodes for in vivo and in vitro electrophysiological signal recording. Adv Healthc Mater 2021;10:e2100646.

3. Eskandarian L, Toossi A, Nassif F, et al. 3D-knit dry electrodes using conductive elastomeric fibers for long-term continuous electrophysiological monitoring. Adv Materials Technologies 2022;7:2101572.

4. Carneiro MR, Majidi C, Tavakoli M. Multi-electrode printed bioelectronic patches for long-term electrophysiological monitoring. Adv Funct Materials 2022;32:2205956.

5. Wang Y, Haick H, Guo S, et al. Skin bioelectronics towards long-term, continuous health monitoring. Chem Soc Rev 2022;51:3759-93.

6. Zhao H, Su R, Teng L, et al. Recent advances in flexible and wearable sensors for monitoring chemical molecules. Nanoscale 2022;14:1653-69.

7. Chen H, Dejace L, Lacour SP. Electronic skins for healthcare monitoring and smart prostheses. Annu Rev Control Robot Auton Syst 2021;4:629-50.

8. Lyu Q, Gong S, Yin J, Dyson JM, Cheng W. Soft wearable healthcare materials and devices. Adv Healthc Mater 2021;10:e2100577.

9. Kim H, Kim E, Choi C, Yeo WH. Advances in soft and dry electrodes for wearable health monitoring devices. Micromachines 2022;13:629.

10. Li Z, Guo W, Huang Y, Zhu K, Yi H, Wu H. On-skin graphene electrodes for large area electrophysiological monitoring and human-machine interfaces. Carbon 2020;164:164-70.

11. Wu H, Yang G, Zhu K, et al. Materials, devices, and systems of on-skin electrodes for electrophysiological monitoring and human-machine interfaces. Adv Sci 2021;8:2001938.

12. Tian G, Yang D, Liang C, et al. A nonswelling hydrogel with regenerable high wet tissue adhesion for bioelectronics. Adv Mater 2023;35:e2212302.

13. Yang L, Liu Q, Zhang Z, Gan L, Zhang Y, Wu J. Materials for dry electrodes for the electroencephalography: advances, challenges, perspectives. Adv Materials Technologies 2022;7:2100612.

14. Liu Q, Yang L, Zhang Z, Yang H, Zhang Y, Wu J. The feature, performance, and prospect of advanced electrodes for electroencephalogram. Biosensors 2023;13:101.

15. Li G, Wang S, Duan YY. Towards conductive-gel-free electrodes: understanding the wet electrode, semi-dry electrode and dry electrode-skin interface impedance using electrochemical impedance spectroscopy fitting. Sensor Actuat B-Chem 2018;277:250-60.

16. Yuan H, Li Y, Yang J, et al. State of the art of non-invasive electrode materials for brain-computer interface. Micromachines 2021;12:1521.

17. Fu Y, Zhao J, Dong Y, Wang X. Dry electrodes for human bioelectrical signal monitoring. Sensors 2020;20:3651.

18. Asl SN, Oehler M, Schilling M. Noise model of capacitive and textile capacitive noncontact electrodes for bioelectric applications. IEEE Trans Biomed Circuits Syst 2018;12:851-9.

19. Ren L, Liu B, Zhou W, Jiang L. A mini review of microneedle array electrode for bio-signal recording: a review. IEEE Sensors J 2020;20:577-90.

20. Wang Y, Jiang L, Ren L, et al. Towards improving the quality of electrophysiological signal recordings by using microneedle electrode arrays. IEEE Trans Biomed Eng 2021;68:3327-35.

21. Yang JC, Mun J, Kwon SY, Park S, Bao Z, Park S. Electronic skin: recent progress and future prospects for skin-attachable devices for health monitoring, robotics, and prosthetics. Adv Mater 2019;31:e1904765.

22. Ren L, Chen Z, Wang H, Dou Z, Liu B, Jiang L. Fabrication of bendable microneedle-array electrode by magnetorheological drawing lithography for electroencephalogram recording. IEEE Trans Instrum Meas 2020;69:8328-34.

23. Huang D, Li J, Li T, Wang Z, Wang Q, Li Z. Recent advances on fabrication of microneedles on the flexible substrate. J Micromech Microeng 2021;31:073001.

24. Li J, Ma Y, Huang D, et al. High-performance flexible microneedle array as a low-impedance surface biopotential dry electrode for wearable electrophysiological recording and polysomnography. Nanomicro Lett 2022;14:132.

25. Hou Y, Li Z, Wang Z, Yu H. Miura-ori structured flexible microneedle array electrode for biosignal recording. Microsyst Nanoeng 2021;7:53.

26. Li Y, Zhou W, Liu C, et al. Fabrication and characteristic of flexible dry bioelectrodes with microstructures inspired by golden margined century plant leaf. Sensor Actuat A-Phys 2021;321:112397.

27. Niu X, Wang L, Li H, Wang T, Liu H, He Y. Fructus xanthii-inspired low dynamic noise dry bioelectrodes for surface monitoring of ECG. ACS Appl Mater Interfaces 2022;14:6028-38.

28. Zhang L, Kumar KS, He H, et al. Fully organic compliant dry electrodes self-adhesive to skin for long-term motion-robust epidermal biopotential monitoring. Nat Commun 2020;11:4683.

29. Cao J, Yang X, Rao J, et al. Stretchable and self-adhesive pedot:pss blend with high sweat tolerance as conformal biopotential dry electrodes. ACS Appl Mater Interfaces 2022;14:39159-71.

30. Park T, Jeong J, Kim YJ, Yoo H. Weak molecular interactions in organic composite dry film lead to degradable, robust wireless electrophysiological signal sensing. Adv Materials Inter 2022;9:2200594.

31. Li Q, Chen G, Cui Y, et al. Highly thermal-wet comfortable and conformal silk-based electrodes for on-skin sensors with sweat tolerance. ACS Nano 2021;15:9955-66.

32. Zhang S, Sharifuzzamn M, Rana SMS, et al. Highly conductive, stretchable, durable, skin-conformal dry electrodes based on thermoplastic elastomer-embedded 3D porous graphene for multifunctional wearable bioelectronics. Nano Res 2023;16:7627-37.

33. Fan W, Zhong Z, Tian G, Wang Y, Gong G, Qi D. Application of Conductive Polymer in Nerve Interface Electrode. Chem J Chin Univ 2021;42:1146-55.

34. Yun G, Tang SY, Sun S, et al. Liquid metal-filled magnetorheological elastomer with positive piezoconductivity. Nat Commun 2019;10:1300.

35. Zhang J, Liu M, Pearce G, et al. Strain stiffening and positive piezoconductive effect of liquid metal/elastomer soft composites. Compos Sci Technol 2021;201:108497.

36. Niu Y, Tian G, Liang C, et al. Thermal-sinterable egain nanoparticle inks for highly deformable bioelectrode arrays. Adv Healthc Mater 2023;12:e2202531.

37. Pei D, Yu S, Liu P, et al. Reversible wet-adhesive and self-healing conductive composite elastomer of liquid metal. Adv Funct Materials 2022;32:2204257.

38. Shi C, Hu F, Wu R, et al. New silk road: from mesoscopic reconstruction/functionalization to flexible meso-electronics/photonics based on cocoon silk materials. Adv Mater 2021;33:e2005910.

39. Hu M, Zhang J, Liu Y, et al. Highly conformal polymers for ambulatory electrophysiological sensing. Macromol Rapid Commun 2022;43:e2200047.

40. Yang H, Ji S, Chaturvedi I, et al. Adhesive biocomposite electrodes on sweaty skin for long-term continuous electrophysiological monitoring. ACS Materials Lett 2020;2:478-84.

41. Meng L, Fu Q, Hao S, Xu F, Yang J. Self-adhesive, biodegradable silk-based dry electrodes for epidermal electrophysiological monitoring. Chem Eng J 2022;427:131999.

42. Zhao Y, Zhang S, Yu T, et al. Ultra-conformal skin electrodes with synergistically enhanced conductivity for long-time and low-motion artifact epidermal electrophysiology. Nat Commun 2021;12:4880.

43. Cheng Y, Zhou Y, Wang R, et al. An elastic and damage-tolerant dry epidermal patch with robust skin adhesion for bioelectronic interfacing. ACS Nano 2022;16:18608-20.

44. Tang W, Zhou Y, Chen S, et al. Delamination-resistant imperceptible bioelectrode for robust electrophysiological signals monitoring. ACS Materials Lett 2021;3:1385-93.

45. Jiang Z, Chen N, Yi Z, et al. A 1.3-micrometre-thick elastic conductor for seamless on-skin and implantable sensors. Nat Electron 2022;5:784-93.

46. Yao S, Zhou W, Hinson R, et al. Ultrasoft porous 3d conductive dry electrodes for electrophysiological sensing and myoelectric control. Adv Mater Technol 2022;7:2101637.

47. Pei Z, Zhang Q, Li Q, et al. A fully 3D printed electronic skin with bionic high resolution and air permeable porous structure. J Colloid Interface Sci 2021;602:452-8.

48. Tian Q, Zhao H, Zhou R, et al. Ultrapermeable and wet-adhesive monolayer porous film for stretchable epidermal electrode. ACS Appl Mater Interfaces 2022;14:52535-43.

49. Xie R, Li Q, Teng L, et al. Strenuous exercise-tolerance stretchable dry electrodes for continuous multi-channel electrophysiological monitoring. npj Flex Electron 2022;6:75.

50. Yu L, Lu L, Zhou X, Xu L. Current understanding of the wettability of mxenes. Adv Materials Inter 2023;10:2201818.

51. Song D, Ye G, Zhao Y, Zhang Y, Hou X, Liu N. An all-in-one, bioderived, air-permeable, and sweat-stable mxene epidermal electrode for muscle theranostics. ACS Nano 2022;16:17168-78.

52. Liu H, Zhong X, He X, et al. Stretchable conductive fabric enabled by surface functionalization of commercial knitted cloth. ACS Appl Mater Interfaces 2021;13:55656-65.

53. Zhang Y, Zhang T, Huang Z, Yang J. A new class of electronic devices based on flexible porous substrates. Adv Sci 2022;9:e2105084.

54. Jeong W, Park Y, Gwon G, et al. All-organic, solution-processed, extremely conformal, mechanically biocompatible, and breathable epidermal electrodes. ACS Appl Mater Interfaces 2021;13:5660-7.

55. Yan X, Chen S, Zhang G, et al. Highly breathable, surface-hydrophobic and wet-adhesive silk based epidermal electrode for long-term electrophysiological monitoring. Compos Sci Technol 2022;230:109751.

56. Zhao Q, Zhu M, Tian G, et al. Highly sensitive and omnidirectionally stretchable bioelectrode arrays for in vivo neural interfacing. Adv Healthc Mater 2023:e2203344.

57. Xing X, Wang Y, Pei W, et al. A high-speed SSVEP-Based BCI using dry EEG electrodes. Sci Rep 2018;8:14708.

58. Liu J, Liu X, He E, et al. A novel dry-contact electrode for measuring electroencephalography signals. Sensor Actuat A-Phys 2019;294:73-80.

59. Niu X, Gao X, Wang T, Wang W, Liu H. Ordered nanopillar arrays of low dynamic noise dry bioelectrodes for electrocardiogram surface monitoring. ACS Appl Mater Interfaces 2022;14:33861-70.

60. Ye G, Qiu J, Fang X, et al. A Lamellibranchia-inspired epidermal electrode for electrophysiology. Mater Horiz 2021;8:1047-57.

61. Dong J, Peng Y, Wang D, et al. Quasi-homogeneous and hierarchical electronic textiles with porosity-hydrophilicity dual-gradient for unidirectional sweat transport, electrophysiological monitoring, and body-temperature visualization. Small 2023;19:e2206572.

62. Stauffer F, Thielen M, Sauter C, et al. Skin conformal polymer electrodes for clinical ecg and eeg recordings. Adv Healthc Mater 2018;7:e1700994.

63. Kim DW, Baik S, Min H, et al. Highly permeable skin patch with conductive hierarchical architectures inspired by amphibians and octopi for omnidirectionally enhanced wet adhesion. Adv Funct Mater 2019;29:1807614.

64. Min H, Baik S, Kim J, et al. Tough carbon nanotube-implanted bioinspired three-dimensional electrical adhesive for isotropically stretchable water-repellent bioelectronics. Adv Funct Materials 2022;32:2107285.

65. Li P, Bao Y, Chen B, et al. A bioinspired sweat-drainable janus electrophysiological electrode for scientific sports training. Adv Materials Technologies 2022;7:2200040.

66. Yang D, Tian G, Liang C, et al. Double-microcrack coupling stretchable neural electrode for electrophysiological communication. Adv Funct Mater 2023;33:2300412.

67. Chen B, Kang W, Sun J, et al. Programmable living assembly of materials by bacterial adhesion. Nat Chem Biol 2022;18:289-94.

68. He K, Liu Z, Wan C, et al. An on-skin electrode with anti-epidermal-surface-lipid function based on a zwitterionic polymer brush. Adv Mater 2020;32:e2001130.

69. Huang Y, Yang F, Liu S, Wang R, Guo J, Ma X. Liquid metal-based epidermal flexible sensor for wireless breath monitoring and diagnosis enabled by highly sensitive SnS2 nanosheets. Research 2021;2021:9847285.

70. Wang X, Feng Z, Zhang G, et al. Flexible sensors array based on frosted microstructured ecoflex film and tpu nanofibers for epidermal pulse wave monitoring. Sensors 2023;23:3717.

71. Zou X, Xue J, Li X, et al. High-fidelity sEMG signals recorded by an on-skin electrode based on AgNWs for hand gesture classification using machine learning. ACS Appl Mater Interfaces 2023;15:19374-83.

72. Dong P, Song Y, Yu S, et al. Electromyogram-based lip-reading via unobtrusive dry electrodes and machine learning methods. Small 2023;19:e2205058.

73. Alizadeh-Meghrazi M, Sidhu G, Jain S, et al. A mass-producible washable smart garment with embedded textile emg electrodes for control of myoelectric prostheses: a pilot study. Sensors 2022;22:666.

74. Cui T, Qiao Y, Li D, et al. Multifunctional, breathable MXene-PU mesh electronic skin for wearable intelligent 12-lead ECG monitoring system. Chem Eng J 2023;455:140690.

75. Wan C, Wu Z, Ren M, et al. In situ formation of conductive epidermal electrodes using a fully integrated flexible system and injectable photocurable ink. ACS Nano 2023;17:10689-700.

76. Zhang X, Mo X, Li C, et al. A wearable master–slave rehabilitation robot based on an epidermal array electrode sleeve and multichannel electromyography network. Adv Intell Syst-Ger 2023;5:2200313.

77. Parak J, Salonen M, Myllymäki T, Korhonen I. Comparison of heart rate monitoring accuracy between chest strap and vest during physical training and implications on training decisions. Sensors 2021;21:8411.

78. Chun S, Kim S, Kim J. Human arm workout classification by arm sleeve device based on machine learning algorithms. Sensors 2023;23:3106.

79. Liu Y, Cheng Y, Shi L, Wang R, Sun J. Breathable, self-adhesive dry electrodes for stable electrophysiological signal monitoring during exercise. ACS Appl Mater Interfaces 2022;14:12812-23.

80. Liang C, Sun J, Liu Z, et al. Wide range strain distributions on the electrode for highly sensitive flexible tactile sensor with low hysteresis. ACS Appl Mater Interfaces 2023;15:15096-107.

Cite This Article

Export citation file: BibTeX | RIS

OAE Style

Liu H, Chen X, Wang Z, Liu Y, Liang C, Zhu M, Qi D. Development of soft dry electrodes: from materials to structure design. Soft Sci 2023;3:27. http://dx.doi.org/10.20517/ss.2023.16

AMA Style

Liu H, Chen X, Wang Z, Liu Y, Liang C, Zhu M, Qi D. Development of soft dry electrodes: from materials to structure design. Soft Science. 2023; 3(3): 27. http://dx.doi.org/10.20517/ss.2023.16

Chicago/Turabian Style

Liu, Hua, Xiaodi Chen, Zihao Wang, Yan Liu, Cuiyuan Liang, Ming Zhu, Dianpeng Qi. 2023. "Development of soft dry electrodes: from materials to structure design" Soft Science. 3, no.3: 27. http://dx.doi.org/10.20517/ss.2023.16

ACS Style

Liu, H.; Chen X.; Wang Z.; Liu Y.; Liang C.; Zhu M.; Qi D. Development of soft dry electrodes: from materials to structure design. Soft. Sci. 2023, 3, 27. http://dx.doi.org/10.20517/ss.2023.16

About This Article

© The Author(s) 2023. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, sharing, adaptation, distribution and reproduction in any medium or format, for any purpose, even commercially, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Data & Comments

Data

Views
633
Downloads
91
Citations
1
Comments
0
0

Comments

Comments must be written in English. Spam, offensive content, impersonation, and private information will not be permitted. If any comment is reported and identified as inappropriate content by OAE staff, the comment will be removed without notice. If you have any queries or need any help, please contact us at support@oaepublish.com.

0
Download PDF
Cite This Article 20 clicks
Like This Article 0 likes
Share This Article
Scan the QR code for reading!
See Updates
Contents
Figures
Related
Soft Science
ISSN 2769-5441 (Online)
Follow Us

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/