Perspective  |  Open Access  |  13 Sep 2023

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Views: 400 |  Downloads: 94 |  Cited:   1
Conn Health Telemed 2023;2:200008.
10.20517/chatmed.2023.03 |  © The Author(s) 2023.
Author Information
Article Notes
Cite This Article

Abstract

The measurement and monitoring of continuous arterial blood pressure (BP) have undergone significant evolution over the past 170 years, transitioning from ancient invasive approaches, like kymograph, to modern non-invasive and unobtrusive technologies such as tonoarteriography (TAG). This progressive shift has revolutionized the way we track BP, providing safer, more accurate, and convenient methods for monitoring BP. This paper aims to provide some historical perspectives on the development of continuous BP technology, highlight key milestones that have shaped the field, discuss the state-of-the-art two-dimensional TAG imaging, and address challenges for future unobtrusive BP measurements. In addition to presenting a concise review of the progression of continuous BP measurement technologies, this article also emphasizes the importance of adopting more precise, convenient and affordable approaches for personalized BP monitoring at home and patient care optimizations at hospitals, thereby empowering healthcare professionals to enhance pervasive hypertension management anywhere.

Keywords

Blood pressure, unobtrusive BP, continuous BP, cuffless BP, two-dimensional tonoarteriographic imaging, hypertension

INTRODUCTION

Hypertension, as the leading preventable risk factor for cardiovascular diseases (CVDs), poses a significant global health challenge. With an estimated 1.3 billion people affected worldwide, it claims the lives of approximately 10 million individuals annually[1]. To effectively prevent and intervene in hypertension and its associated CVDs, reliable and timely measurement, diagnosis, and monitoring of BP are crucial[2]. While conventional office BP measurement is the established gold standard[3], it provides only a single assessment under conditions that can influence the level of BP being measured[4]. Recognizing this limitation, out-of-office BP measurements, including home BP monitoring (HBPM) and ambulatory BP monitoring (ABPM), have gained prominence for their additional diagnostic value in hypertension care interventions[5]. Specifically, patient-initiated HBPM is increasingly preferred due to improved BP control, diagnosis of white-coat hypertension, telemonitoring feasibility, and cardiovascular risk prediction[6]. Moreover, HBPM is also more cost-effective and efficient compared to ABPM, particularly for long-term hypertension screening[7,8]. The significance of personalized long-term BP monitoring at home is widely recognized[9], but commonly used methods, such as auscultatory and oscillometric, have limitations that hinder their effectiveness in this context. Specifically, the use of an inflatable cuff causes discomfort among individuals with hypertension, necessitating frequent readings, and it lacks the ability to perform nocturnal recording, which is critical in preventing sudden deaths[10]. Additionally, these methods only provide intermittent measurements and cannot adequately monitor short- or long-term BP changes, which hold importance in assessing individuals at risk who may not conventionally be considered candidates for treatment benefits[11-14].

Challenges to the implementation of out-of-office BP measurement were patient inertia, poor patient compliance, lack of medical consultation time, limited access to a sphygmomanometer and shortage of related educational materials[5]. Recognizing the indispensability of home-based long-term continuous BP monitoring, a range of strategies are being explored to improve BP control, including individualized treatment, behavioral interventions, user-friend devices, mobile health apps, and online platforms, all of which underscore the essential role of treatment adherence and consistency[15].

In light of these, the purpose of this paper is to provide a concise review of the historical evolutions of continuous BP measurement technology, tracing its evolution from traditional invasive methods to the current advanced TAG techniques[16]. By examining key milestones and advancements in the field, this review offers insights into the challenges faced in BP monitoring and highlights potential opportunities for future developments.

MILESTONES OF CONTINUOUS BP MEASUREMENT TECHNIQUES

This section offers an overview of the progression of continuous BP measurements, categorized into four stages, including the development of invasive methods, non-invasive approaches, the emergence of cuffless technologies, and the current frontiers of TAG imaging, as shown in Figure 1.

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Figure 1. Diagram of four stages in the evolution of continuous blood pressure (BP) measurements.

Developments in invasive BP measurement approaches

The history of continuous BP measurement dates back to the emergence of invasive BP techniques in 1733, when Stephen Hales observed fluctuating blood levels in a glass tube inserted into a horse's artery, laying the foundation for future developments[17]. A century later, a significant achievement occurred in 1847 with Ludwig's development of the kymograph, which enabled graphical recording of continuous oscillations of human arterial pressure through arterial cannulation, revolutionizing the acquisition of valuable hemodynamic information[18]. In 1949, intra-arterial continuous BP measurement through cannulation was first implemented in clinical settings[19]. Subsequently, technological and medical advancements during the 20th century propelled the further development of invasive BP measurement techniques, including the refinement of catheterization methods, introduction of electronic amplifiers and transducers, and adoption of disposable catheters, significantly enhancing reliability, real-time monitoring, and availability of invasive BP monitoring. Today, invasive BP (IBP) method is considered the gold standard in ensuring accurate and continuous BP measurement in intensive care units (ICU), operating rooms, and other clinical environments, with the operational diagram depicted in Figure 2[20]. Based on empirical clinical data, this approach demonstrates a commendable reliability and stability, as evidenced by Mean Error (ME) of 3 mmHg and Standard Deviation (SD) of 2 mmHg[21-23]. Nevertheless, it is essential for physicians to be mindful of the potential accuracy limitations of IBP in certain patients due to underdamping/resonance phenomena, as highlighted by a study revealing underdamping/resonance occurrence in 30.7% (92 out of 300 subjects) of arterial signals[24]. Additionally, this approach is not without inherent risks, including symptomatic or asymptomatic arterial thrombosis, infections, accidental injection of intravenous drugs, and other discomforts associated with catheterization, rendering continuous monitoring in daily activities impractical[25]. Consequently, the demand for non-invasive BP measurements has become an ongoing challenge, prompting the necessity for further enhancements in the field.

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Figure 2. Operational diagram of clinical invasive blood pressure (BP) monitoring device.

Advancements in non-invasive continuous BP techniques

In contrast to invasive methods, non-invasive BP (NIBP) measurement techniques offer improved safety, reproducibility, and procedural simplicity[26]. The exploration of NIBP began in 1963 when G.L. Pressman and P.M. Newgard introduced the first arterial tonometer, accompanied by the development of a discrete, linear mechanical model[27]. Another significant development occurred in 1973 when Peñáz devised the volume clamp method[28], as shown in Figure 3. These techniques continue to be widely used in continuous BP monitoring today and have played an important role in the widespread adoption of home monitoring. There are increasing studies to assess the accuracy of NIBP methods. For example, Saugel et al. examined the feasibility of arterial pressure measurement using tonometer technology in 34 ICU patients[29]. They compared TL-200pro (Tensys Medical, Inc., San Diego, CA, USA) radial artery readings to femoral catheter values, revealing bias and limits of agreement for Mean Arterial Pressure (MAP) (+0.72 mmHg, -9.37 to +10.82 mmHg), Systolic Arterial Pressure (SAP) (-1.39 mmHg, -18.74 to +15.96 mmHg), and Diastolic Arterial Pressure (DAP) (+4.36 mmHg, -8.66 to +17.38 mmHg) from 4,502 averaged 10-beat epochs, with errors at 12%, 14%, and 21%[29]. Moreover, Jagadeesh et al. found 94.5%, 95.1%, and 99.4% agreement for Systolic Blood Pressure (SBP), Diastolic Blood Pressure (DBP), and MAP, respectively, demonstrating volume clamp technology's reliability comparable to invasive methods[30]. While NIBP methods show promise in terms of accuracy, they are facing challenges including unwieldy design and discomfort caused by cuffs, necessitating external force. At the same time, practical issues on optimal sensor positioning, susceptibility to motion-induced artifacts and complex calibration remain significant considerations[31,32], limiting their suitability for continuous daily use. Thus, substantial efforts are needed to attain ubiquitous, unobtrusive, and continuous BP monitoring.

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Figure 3. Product forms and block diagrams for (A) arterial tonometer and (B) volume clamp system[11].

Achievements in continuous cuffless BP technologies

Recent technological improvements have made notable strides in the field of continuous BP measurement, particularly in the development of cuffless techniques. Different from traditional mechanical-based BP measuring methods such as arterial line and stethoscope, cuffless measurement methods commonly employ techniques such as photoplethysmography (PPG), electrocardiography (ECG), impedance plethysmography (IPG) and innovative camera-based remote PPG[33], which eliminate the need for traditional cuffs, providing more convenient and pervasive methods for measuring BP.

Over the past two decades, significant achievements have been made in the development of wearable cuffless BP/TAG measuring devices, allowing for continuous BP monitoring in daily life. Notably, Li et al. have contributed to the early development of various cuffless BP measurement models and platforms[34], as depicted in Figure 4, including BP ring[35], BP watch[36], h-Shirt[37], BP mobile phones[38], u-Chair[39], u-Bed [40], BP glasses[41], BP armband[42], etc., which utilize both physiological mechanism model and data-driven or AI model-based methods for BP estimation. Remarkably, they have introduced multi-wavelength photoplethysmography (MWPPG) technology[43,44], which has been validated by Liu et al. in a large-scale study involving 3,077 participants, demonstrating the reasonable performance of the MWPPG-BP method[45]. Overall, these remarkable innovations in wearable continuous cuffless BP measurement devices, coupled with extensive research and validation studies conducted by multiple teams, have offered the possibility of realizing convenient, unobtrusive, continuous monitoring of BP without inflation/deflation cuff for daily uses.

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Figure 4. Some major prototype developments of continuous cuffless blood pressure (BP)/tonoarteriography (TAG) measuring devices over the past 20 years.

However, cuffless BP devices are not recommended for clinical use according to the 2021 European Society of Hypertension Guidelines[46], possibly due to ongoing concerns about their accuracy. These concerns arise from factors such as validation problems, issues with sensor placement, potential motion artifacts, and calibration challenges. While many studies claim to adhere to standards for BP measurement accuracy, such as the AAMI/ESH/ISO Universal Standard (imposing bias and precision errors within 5 and 8 mmHg) or the IEEE Std 1708TM standard for wearable cuffless BP devices [requiring a mean absolute difference (MAD) ≤ 7 mmHg][47], it is important to note that these claims might not fully follow all the essential criteria outlined by these standards[46,48-51]. However, there are some approaches for cuffless BP estimation that have demonstrated promising outcomes. For instance, a pioneering algorithm integrating PTT and photoplethysmogram intensity ratio (PIR) conducted the validation on a cohort of 27 subjects, resulting in mean ± SD values of -0.37 ± 5.21 mmHg, -0.08 ± 4.06 mmHg and -0.18 ± 4.13 mmHg for systolic, diastolic, and mean BP respectively, with corresponding MAD measuring 4.09 mmHg, 3.18 mmHg and 3.18 mmHg. Remarkably, their findings exceeded the performance of the two most cited PTT algorithms by approximately 2 mmHg in both SD and MAD[52]. This underscores a positive trajectory in the advancement of cuffless BP measurement techniques.

Two-dimensional (2D) TAG imaging technologies

The aforementioned BP measurement techniques primarily focus on single-point or one-dimensional time serial signals, limiting the ability to comprehensively assess BP distribution across different regions[53]. Recognizing the substantial influence of measurement sites on BP evaluation, we have proposed a novel approach known as TAG imaging system, a cost-effective, wearable and non-invasive technique used for continuous 2D BP monitoring based on a homemade electronic-optical sensor array[54].

The architecture of the TAG imaging system is graphically depicted in Figure 5. Comprising a wearable multichannel signal acquisition configuration, this apparatus features an arrangement of 9 PPG sensors. These sensors are strategically positioned to encompass the radial and ulnar arteries, as well as the adjacent regions, optimizing spatial coverage. Imbedded within a designed wristband, these sensors wrapped by the elastic band hold a close and conforming interface with the skin. Additionally, we employed a 16-channel data collection module (manufactured by Biopac Systems Inc., USA) to gather data simultaneously and continuously for the analysis and mapping of PPG, PTT, and BP.

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Figure 5. Architecture of the tonoarteriography (TAG) imaging visualization system.

This system visualizes the distribution of PPG/BP at the wrist using a multi-channel PPG array and one-lead ECG. BP is derived based on the PIR-PTT equations (1-2), as introduced by Ding et al.[52,55]:

$$ D B P=D B P_{0} \cdot \frac{P I R_{0}}{P I R} $$

$$ S B P=D B P_{0} \cdot \frac{P I R_{0}}{P I R}+P P_{0} \cdot\left(\frac{P T T_{0}}{P T T}\right)^{2} $$

where PTT is derived from calculations on ECG and localized PPG signals, DBP0, PIR0, PP0 and PTT0 are calibrated from continuous BP from Biopac. Distinct variations in local peripheral BP can be identified, providing valuable insights into BP distribution across different regions. The estimated differences in SBP and DBP from different channels are calculated using the equation (3):

$$ \text { Difference }=B P_{\text {Reference }}-B P_{C h X} $$

where BPReference are continuous BPs obtained from Biopac, and BPChx are estimated BPs from Channel X (X = 1, 2, 3…9). Subsequently, these variations are graphically represented through heatmap and color fill, as illustrated in Figure 6. The study presents a two-dimensional framework for regional BP measurement, offering potential applications in evaluating peripheral vascular conditions and improving central or brachial BP estimation accuracy for effective hypertension management.

Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging

Figure 6. Mapping of estimated difference distribution of blood pressure (BP) values from different channels[54].

Compared with the current commercial BP measurement products, the proposed system offers a potential solution to addressing issues related to inconsistent device placement, which is the limitation inherent in single-point cuffless BP measurements. It holds promise for delving into the complex relationship between signal accuracy and spatial positioning, leading to improved precision and a wider range of physiological parameter estimations within confined measurement locations. Moreover, the system demonstrates the capability to improve the accuracy of central or brachial BP estimations, thereby providing substantial support for clinical decision-making. Beyond this, the novel non-invasive and cuffless imaging approach overcomes the constraints posed by traditional inflatable cuff-based methods, not only enhancing patients’ comfort, compliance and continuous BP monitoring but also proving additional local geographic-dependent BP information useful for studying clinical micro-circulation.

Furthermore, given the ongoing concerns regarding accuracy, the validation methodology for the proposed BP monitoring system is designed to closely adhere to the scenario-based validation test protocols stipulated by the European Society of Hypertension (ESH) for the Validation of Cuffless Blood Pressure Measuring Devices[56]. The initial phase involves static tests for quantifying the absolute precision of BP measurements. Additionally, compliance with the IEEE Std 1708TM standard, which is a landmark contribution in the era of cuffless BP devices[47], serves to reinforce the assurance of precision and reliability in real-world clinical scenarios, thereby contributing to the effective control and treatment of hypertension.

CONCLUSION

The historical evolution of continuous BP measurements has demonstrated a remarkable progression, spanning from the development of invasive techniques to state-of-the-art obtrusive 2D BP/TAG imaging devices, and culminating in the diverse ranges of modern unobtrusive methods, which highlights the growing significance of continuous BP measurement methods in monitoring hypertension[53]. As technology continues to advance, the shift of BP measurement techniques has emphasized the importance of high-dimensional approaches in enhancing accuracy and overcoming the limitations of single-point or snapshot measurements. The assessment of regional BP distribution and evaluation of vascular conditions have emerged as crucial factors for future BP monitoring devices. Notably, the recent emergence of TAG imaging technology holds substantial potential in enriching the evaluation of vascular conditions through visualizing BP distribution at different locations, thereby enhancing the capabilities of early monitoring for peripheral vascular conditions. Looking ahead, we anticipate that further progress in BP measurement techniques will be characterized by multi-modal, multi-parameter, and multi-dimensional approaches. This trajectory will lead to the development of more unobtrusive, intelligent and multifunctional connected wearable devices, resulting in improved accuracy, enhanced usability, and elevated standards of personalized patient care at both home and hospital, ultimately revolutionizing hypertension management in the future.

DECLARATIONS

Authors’ contributions

Made a substantial contribution to the conception and writing of the manuscript: Liu ZJ, Zhang YT

Contributed to the original idea of blood pressure imaging: Zhang YT

Contributed to the manuscript's editing: Liu ZJ, Xiang T, Ji N, Zhang YT

Availability of data and materials

Not applicable.

Financial support and sponsorship

This work was supported by the ITC-InnoHK grant to Hong Kong Center for Cerebro-cardiovascular Health Engineering (COCHE) and City University of Hong Kong (CityU) starting up grant.

Conflicts of interest

All authors declared that there are no conflicts of interest.

Ethical approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Copyright

© The Author(s) 2023.

REFERENCES

1. World hypertension day: taking action against the silent epidemic of high blood pressure. Available from: https://world-heart-federation.org/news/world-hypertension-day-taking-action-against-the-silent-epidemic-of-high-blood-pressure/ [Last accessed on 7 Sep 2023].

2. Ringrose JS, Padwal R. The current state of blood pressure measurement and emerging technologies. Conn Health Telemed 2023;2:200002.

3. Mancia G, Fagard R, Narkiewicz K, et al. Task Force for the Management of Arterial Hypertension of the European Society of Hypertension and the European Society of Cardiology. 2013 ESH/ESC practice guidelines for the management of arterial hypertension. Blood Press 2014;23:3-16.

4. O'Brien E, Parati G, Stergiou G, et al. European Society of Hypertension Working Group on Blood Pressure Monitoring. European Society of hypertension position paper on ambulatory blood pressure monitoring. J Hypertens 2013;31:1731-68.

5. Setia S, Subramaniam K, Teo BW, Tay JC. Ambulatory and home blood pressure monitoring: gaps between clinical guidelines and clinical practice in Singapore. Int J Gen Med 2017;10:189-97.

6. George J, MacDonald T. Home blood pressure monitoring. Eur Cardiol 2015;10:95-101.

7. Imai Y, Obara T, Asamaya K, Ohkubo T. The reason why home blood pressure measurements are preferred over clinic or ambulatory blood pressure in Japan. Hypertens Res 2013;36:661-72.

8. Arrieta A, Woods JR, Qiao N, Jay SJ. Cost-benefit analysis of home blood pressure monitoring in hypertension diagnosis and treatment: an insurer perspective. Hypertension 2014;64:891-6.

9. Jaeger BC, Bress AP, Bundy JD, et al. Longer-term all-cause and cardiovascular mortality with intensive blood pressure control: a secondary analysis of a randomized clinical trial. JAMA Cardiol 2022;7:1138-46.

10. Plange N, Kaup M, Daneljan L, Predel HG, Remky A, Arend O. 24-h blood pressure monitoring in normal tension glaucoma: night-time blood pressure variability. J Hum Hypertens 2006;20:137-42.

11. Ding XR, Zhao N, Yang GZ, et al. Continuous blood pressure measurement from invasive to unobtrusive: celebration of 200th birth anniversary of carl ludwig. IEEE J Biomed Health Inform 2016;20:1455-65.

12. Schutte AE, Kollias A, Stergiou GS. Blood pressure and its variability: classic and novel measurement techniques. Nat Rev Cardiol 2022;19:643-54.

13. Dai Y, Wang Y, Xie Y, et al. Short-term and long-term blood pressure changes and the risk of all-cause and cardiovascular mortality. Biomed Res Int 2019;2019:5274097.

14. Mamun MMR, Sherif A. Advancement in the cuffless and noninvasive measurement of blood pressure: a review of the literature and open challenges. Bioengineering 2022;10:27.

15. Mancia G, Cappuccio FP, Burnier M, et al. Perspectives on improving blood pressure control to reduce the clinical and economic burden of hypertension. J Intern Med 2023;294:251-68.

16. Ding X, Dai W, Luo N, Liu J, Zhao N, Zhang Y. A flexible tonoarteriography-based body sensor network for cuffless measurement of arterial blood pressure. 2015 IEEE 12th international conference on wearable and implantable body sensor networks (BSN). Cambridge, USA; 2015. pp. 1-4.

17. Smith IB. The impact of Stephen Hales on medicine. J R Soc Med 1993;86:349-52.

18. Valentinuzzi ME, Beneke K, González GE. Ludwig: the bioengineer. IEEE Pulse 2012;3:68-9, 72, 74.

19. Peterson LH, Dripps RD, Risman GC. A method for recording the arterial pressure pulse and blood pressure in man. Am Heart J 1949;37:771-82.

20. Pascual JL, Horak J, Gracias VH, Neligan PJ. Chapter 19-volume status and cardiac function. Monitoring in neurocritical care. Elsevier; 2013. pp. 176-88.e3.

21. Rook WH, Turner JD, Clutton-Brock TH. Analysis of damping characteristics of arterial catheter blood pressure monitoring in a large intensive care unit. South Afr J Crit Car 2017;33:8-10. Available from: https://www.ajol.info/index.php/sajcc/article/view/159671 [Last accessed on 7 Sep 2023]

22. Yeomanson CW, Evans DH. The frequency response of external transducer blood pressure measurement systems: a theoretical and experimental study. Clin Phys Physiol Meas 1983;4:435-49.

23. Lam S, Liu H, Jian Z, Settels J, Bohringer C. Intraoperative invasive blood pressure monitoring and the potential pitfalls of invasively measured systolic blood pressure. Cureus 2021;13:e17610.

24. Romagnoli S, Ricci Z, Quattrone D, et al. Accuracy of invasive arterial pressure monitoring in cardiovascular patients: an observational study. Crit Care 2014;18:644.

25. Alexander B, Cannesson M, Quill TJ. Chapter 12-Blood pressure monitoring. Anesthesia equipment. Elsevier; 2013. pp. 273-82.

26. Ismail SNA, Nayan NA, Jaafar R, May Z. Recent advances in non-invasive blood pressure monitoring and prediction using a machine learning approach. Sensors 2022;22:6195.

27. Pressman GL, Newgard PM. A transducer for the continuous external measurement of arterial blood pressure. IEEE Trans Biomed Eng 1963;10:73-81.

28. Penaz J. Photoelectric measurement of blood pressure, volume and flow in the finger. In: digest of the 10th international conference on medical and biological engineering (in Japanese). Available from: https://cir.nii.ac.jp/crid/1570854174382447616 [Last accessed on 7 Sep 2023]

29. Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W. Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth 2013;111:185-90.

30. Jagadeesh AM, Singh NG, Mahankali S. A comparison of a continuous noninvasive arterial pressure (CNAP™) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU. Ann Card Anaesth 2012;15:180-4.

31. Drzewiecki GM, Melbin J, Noordergraaf A. Arterial tonometry: review and analysis. J Biomech 1983;16:141-52.

32. Wu H, Ji Z, Li M. Non-invasive continuous blood-pressure monitoring models based on photoplethysmography and electrocardiography. Sensors 2019;19:5543.

33. Curran T, McDuff D, Liu X, et al. Camera-based remote photoplethysmography for blood pressure measurement: current evidence, clinical perspectives, and future applications. Conn Health Telemed 2023;2:200004.

34. Li S, Zhang C, Xu Z, et al. Cuffless Blood Pressure Monitoring: Academic Insights and Perspectives Analysis. Micromachines 2022;13:1225.

35. Zheng D, Zhang Y. A ring-type device for the noninvasive measurement of arterial blood pressure. Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No. 03CH37439). Cancun, Mexico; 2003. pp. 3184-87.

36. Poon CC, Wong YM, Zhang YT. M-health: the development of cuff-less and wearable blood pressure meters for use in body sensor networks. 2006 IEEE/NLM Life Science Systems and Applications Workshop. Bethesda, USA; 2006, pp. 1-2.

37. Zhang YT, Poon CCY, Chan C, Tsang MWW, Wu K. A health-shirt using e-textile materials for the continuous and cuffless monitoring of arterial blood pressure. 2006 3rd IEEE/EMBS International Summer School on Medical Devices and Biosensors, Cambridge, USA; 2006, pp. 86-89,

38. Poon C, Zhang YT, Bao S. A novel biometrics method to secure wireless body area sensor networks for telemedicine and m-health. IEEE Commun Mag 2006;44:73-81.

39. Wu K, Chan C, Zhang YT. Contactless and cuffless monitoring of blood pressure on a chair using e-textile materials. 2006 3rd IEEE/EMBS International Summer School on Medical Devices and Biosensors. Cambridge, USA; 2006. pp. 98-100.

40. Gu WB, Poon CC, Leung HK, Sy MY, Wong MY, Zhang YT. A novel method for the contactless and continuous measurement of arterial blood pressure on a sleeping bed. Annu Int Conf IEEE Eng Med Biol Soc 2009;2009:6084-6.

41. Zheng Y, Leung B, Sy S, Zhang Y, Poon CC. A clip-free eyeglasses-based wearable monitoring device for measuring photoplethysmograhic signals. Annu Int Conf IEEE Eng Med Biol Soc 2012;2012:5022-5.

42. Zheng YL, Yan BP, Zhang YT, Poon CC. An armband wearable device for overnight and cuff-less blood pressure measurement. IEEE Trans Biomed Eng 2014;61:2179-86.

43. Liu J, Li Y, Ding XR, Dai WX, Zhang YT. Effects of cuff inflation and deflation on pulse transit time measured from ECG and multi-wavelength PPG. Annu Int Conf IEEE Eng Med Biol Soc 2015;2015:5973-6.

44. Liu J, Yan BP, Zhang YT, Ding XR, Su P, Zhao N. Multi-wavelength photoplethysmography enabling continuous blood pressure measurement with compact wearable electronics. IEEE Trans Biomed Eng 2019;66:1514-25.

45. Liu ZD, Li Y, Zhang YT, et al. Cuffless blood pressure measurement using smartwatches: a large-scale validation study. IEEE J Biomed Health Inform 2023;27:4216-27.

46. Mukkamala R, Yavarimanesh M, Natarajan K, et al. Evaluation of the accuracy of cuffless blood pressure measurement devices: challenges and proposals. Hypertension 2021;78:1161-67.

47. IEEE standard for wearable, cuffless blood pressure measuring devices-amendment 1. IEEE Std 1708a-2019 (Amendment to IEEE Std 1708-2014);2019.pp.1-35.

48. Watanabe N, Bando YK, Kawachi T, et al. Development and validation of a novel cuff-less blood pressure monitoring device. JACC Basic Transl Sci 2017;2:631-42.

49. Nachman D, Gepner Y, Goldstein N, et al. Comparing blood pressure measurements between a photoplethysmography-based and a standard cuff-based manometry device. Sci Rep 2020;10:16116.

50. Schoettker P, Degott J, Hofmann G, et al. Blood pressure measurements with the OptiBP smartphone app validated against reference auscultatory measurements. Sci Rep 2020;10:17827.

51. Liu Z, Zhou B, Li Y, Tang M, Miao F. Continuous blood pressure estimation from electrocardiogram and photoplethysmogram during arrhythmias. Front Physiol 2020;11:575407.

52. Ding XR, Zhang YT, Liu J, Dai WX, Tsang HK. Continuous cuffless blood pressure estimation using pulse transit time and photoplethysmogram intensity ratio. IEEE Trans Biomed Eng 2016;63:964-72.

53. Zheng YL, Ding XR, Poon CC, et al. Unobtrusive sensing and wearable devices for health informatics. IEEE Trans Biomed Eng 2014;61:1538-54.

54. Liu ZJ, Xiang T, Ji N, Zhang YT. Optical sensor array-based multi-channel wearable tonoarteriogram (TAG) imaging system. 45th Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2023.

55. Sharma M, Barbosa K, Ho V, et al. Cuff-less and continuous blood pressure monitoring: a methodological review. Technologies 2017;5:21.

56. Stergiou GS, Avolio AP, Palatini P, et al. European society of hypertension recommendations for the validation of cuffless blood pressure measuring devices: european society of hypertension working group on blood pressure monitoring and cardiovascular variability. J Hypertens 2023:online ahead of print.

Cite This Article

Export citation file: BibTeX | RIS

OAE Style

Liu ZJ, Xiang T, Ji N, Zhang YT. Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging. Conn Health Telemed 2023;2:200008. http://dx.doi.org/10.20517/chatmed.2023.03

AMA Style

Liu ZJ, Xiang T, Ji N, Zhang YT. Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging. Connected Health And Telemedicine. 2023; 2(3): 200008. http://dx.doi.org/10.20517/chatmed.2023.03

Chicago/Turabian Style

Liu, Zi-Jun, Ting Xiang, Nan Ji, Yuan-Ting Zhang. 2023. "Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging" Connected Health And Telemedicine. 2, no.3: 200008. http://dx.doi.org/10.20517/chatmed.2023.03

ACS Style

Liu, Z.J.; Xiang T.; Ji N.; Zhang Y.T. Some perspectives of continuous arterial blood pressure measurements: from kymograph to tonoarteriographic imaging. Conn. Health. Telemed. 2023, 2, 200008. http://dx.doi.org/10.20517/chatmed.2023.03

About This Article

© The Author(s) 2023. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, sharing, adaptation, distribution and reproduction in any medium or format, for any purpose, even commercially, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Data & Comments

Data

Views
400
Downloads
94
Citations
1
Comments
0
1

Comments

Comments must be written in English. Spam, offensive content, impersonation, and private information will not be permitted. If any comment is reported and identified as inappropriate content by OAE staff, the comment will be removed without notice. If you have any queries or need any help, please contact us at support@oaepublish.com.

0
Download PDF
Cite This Article 13 clicks
Like This Article 1 likes
Share This Article
Scan the QR code for reading!
See Updates
Contents
Figures
Related
Connected Health And Telemedicine
ISSN 2993-2920 (Online)

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/

Portico

All published articles are preserved here permanently:

https://www.portico.org/publishers/oae/